Skip to content
Menu

Beamex MC6

Technical Specifications

Home » Calibrators » Beamex MC6 » Beamex MC6 » Beamex MC6 Technical Specifications

GENERAL SPECIFICATIONS

FEATURE

VALUE

Display

5.7” Diagonal 640 x 480 TFT LCD module

Touch panel

5-wire resistive touch screen

Keyboard

Membrane keyboard

Backlight

LED backlight, adjustable brightness

Weight

Extended case: 1.5…2.0 kg (3.3…4.4 lb) Flat case: 1.5 kg (3.3 lb)

Dimensions

Extended case: 200 mm × 230 mm × 70 mm (D × W × H)

( 7.87 in × 9.06 in × 2.76 in)

Battery type

Rechargeable lithium-ion polymer, 4200 mAh, 11.1V

Charging time

Approximately 4 hours

Charger supply

100…240 VAC, 50–60 Hz

Battery operation

10…16 hours

Operating temperature

−10…45 °C (14…113 °F)

Operating temperature while charging batteries

0…30 °C (32…86 °F)

Storage temperature

−20…60 °C (−4…140 °F)

Specifications valid

−10…45 °C, unless other mentioned

Humidity

0…80% R.H. non condensing

Warmup time

Specifications valid after a 5 minute warmup period.

Max. input voltage

30 V AC, 60 V DC

Display update rate

3 readings/second

Safety

Directive 2014/35/EU, EN 61010-1:2010 + A1:2019 + AC:2019-04

EMC

Directive 2014/30/EU, EN IEC 61326-1:2021

Ingress protection

IP65

RoHS compliance

ROHS II Directive 2011/65/EU, EN IEC 63000:2018

Drop

IEC 60068-2-32. 1 meter (3.28 ft)

Vibration

IEC 60068-2-64. Random, 2 g, 5…500 Hz

Max altitude

3,000 m (9,842 ft)

Warranty

Warranty 3 years. 1 year for battery pack.

PRESSURE MEASUREMENT

INTERNAL

MODULES

EXTERNAL

MODULES

UNIT

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

PB

EXT B

kPa a
mbar a
psi a

70 … 120
700 … 1200
10.15 … 17.4

0.01
0.1
0.001

0.03 kPa
0.3 mbar
0.0044 psi

0.05 kPa
0.5 mbar
0.0073 psi

P10mD

EXT10mD

kPa

mbar

iwc

±1

±10

±4

0.0001
0.001
0.001

0.05 % Span

0.05 % Span + 0.1 % RDG

P100m

EXT100m

kPa

mbar

iwc

0 … 10

0 … 100

0 … 40

0.0001
0.001
0.001

0.015 % FS + 0.0125 % RDG

0.025 % FS + 0.025 % RDG

P400mC

EXT400mC

kPa

mbar

iwc

±40
±400
±160

0.001
0.01
0.001

0.01 % FS + 0.0125 % RDG

0.02 % FS + 0.025 % RDG

P1C

EXT1C

kPa

bar

psi

c

±100
±1
–14.5 … 15

0.001
0.00001
0.0001

0.007 % FS + 0.0125 % RDG

0.015 % FS + 0.025 % RDG

P2C

EXT2C

kPa

bar

psi

–100 … 200
–1 … 2
–14.5 … 30

0.001
0.00001
0.0001

0.005 % FS + 0.01 % RDG

0.01 % FS + 0.025 % RDG

P6C

EXT6C

kPa

bar

psi

–100 … 600
–1 … 6
–14.5 … 90

0.01
0.0001
0.001

0.005 % FS + 0.01 % RDG

0.01 % FS + 0.025 % RDG

P20C

EXT20C

kPa

bar

psi

–100 … 2000
–1 … 20
–14.5 … 300

0.01
0.0001
0.001

0.005 % FS + 0.01 % RDG

0.01 % FS + 0.025 % RDG

P60

EXT60

kPa

bar

psi

0 … 6000
0 … 60
0 … 900

0.1
0.001
0.01

0.005 % FS + 0.0125 % RDG

0.01 % FS + 0.025 % RDG

P100

EXT100

MPa

bar

psi

0 … 10
0 … 100
0 … 1500

0.0001
0.001
0.01

0.005 % FS + 0.0125 % RDG

0.01 % FS + 0.025 % RDG

P160

EXT160

MPa

bar

psi

0 … 16
0 … 160
0 … 2400

0.0001
0.001
0.01

0.005 % FS + 0.0125 % RDG

0.01 % FS + 0.025 % RDG

EXT250

MPa

bar

psi

0 … 25
0 … 250
0 … 3700

0.001
0.01
0.1

0.007 % FS + 0.0125 % RDG

0.015 % FS + 0.025 % RDG

EXT600

MPa

bar

psi

0 … 60
0 … 600
0 … 9000

0.001
0.01
0.1

0.007 % FS + 0.01 % RDG

0.015 % FS + 0.025 % RDG

EXT1000

MPa

bar

psi

0 … 100
0 … 1000
0 … 15000

0.001
0.01
0.1

0.007 % FS + 0.01 % RDG

0.015 % FS + 0.025 % RDG

Footnotes

1) Accuracy includes hysteresis, nonlinearity and repeatability (k=2).
2) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned period (k=2).
3) Every internal/external gauge pressure module’s range may be displayed also in absolute pressure if the barometric module (PB or EXT B) is installed/connected.

Maximum number of internal pressure modules is 3 gauge/differential pressure modules and one barometric (PB) module.
There is a connection for external pressure modules.
% RDG = Percentage of the actual pressure reading.
% FS = Percentage of the full scale range from zero to maximum.
% Span = Percentage of the total measurement range from minimum to maximum.
SUPPORTED PRESSURE UNITS
Pa, kPa, hPa, MPa, mbar, bar, gf/cm2, kgf/cm2, kgf/m2, kp/cm2, lbf/ft2, psi,at, torr, atm, ozf/in2, iwc, inH2O, ftH2O, mmH2O, cmH2O, mH2O, mmHg, cmHg, mHg, inHg, mmHg(0 °C), inHg(0 °C), mmH2O(60°F), mmH2O(68°F), mmH2O(4 °C), cmH2O(60°F), cmH2O(68°F), cmH2O(4 °C), inH2O(60°F), inH2O(68°F), inH2O(4 °C), ftH2O(60°F), ftH2O(68°F), ftH2O(4 °C).
Large number of user pressure units can be created.
TEMPERATURE COEFFICIENT
<±0.001 % RDG / °C outside 15–35 °C (59–95 °F).
P10mD / EXT10mD: < ±0.002 % Span / °C outside 15–35 °C (59–95 °F).
MAX OVERPRESSURE
2 times the nominal pressure. Except the following modules;
PB/EXTB: 1200 mbar abs (35.4 inHg abs).
P10mD/EXT10mD: 200 mbar (80 iwc).
EXT600: 900 bar (13000 psi).
EXT1000: 1000 bar (15000 psi).
PRESSURE MEDIA
Modules up to P6C/EXT6C: dry, clean air or other clean, inert, non-toxic, noncorrosive gases.
Modules P20C/EXT20C and higher: clean, inert, non-toxic, non-corrosive gases or liquids.
WETTED PARTS
AISI316 stainless steel, Hastelloy, Nitrile rubber.
PRESSURE CONNECTIONS
PB/EXTB: 10/32″ (M5) female.
P10mD/EXT10mD: 2 x adapters for 1/8” ID hose (3,2 mm).
P100m/EXT100m to P20C/EXT20C: Bx G1/8” male compatible with Beamex 40 bar hoses.
P60, P100, P160: Bx 1215 male compatible with Beamex 630 bar hoses.
EXT60 to EXT1000: G1/4″ B male.

THERMOCOUPLE MEASUREMENT & SIMULATION

TC1 measurement & simulation / TC2 measurement

TYPE

RANGE (°C)

RANGE (°C)

ACCURACY (1

1 YEAR UNCERTAINTY (2

B (3

0 … 1820

0 … 200
200 … 500
500 … 800
800 … 1820

(8
1.5 °C
0.6 °C
0.4 °C

(4
2.0 °C
0.8 °C
0.5 °C

R (3

–50 … 1768

–50 … 0
0 … 150
150 … 400
400 … 1768

0.8 °C
0.6 °C
0.35 °C
0.3 °C

1.0 °C
0.7 °C
0.45 °C
0.4 °C

S (3

–50 … 1768

–50 … 0
0 … 100
100 … 300
300 … 1768

0.7 °C
0.6 °C
0.4 °C
0.35 °C

0.9 °C
0.7 °C
0.55 °C
0.45 °C

E (3

–270 … 1000

–270 … –200
–200 … 0
0 … 1000

(8
0.05 °C + 0.04 % RDG
0.05 °C + 0.003 % RDG

(4
0.07 °C + 0.06 % RDG
0.07 °C + 0.005 % RDG

J (3

–210 … 1200

–210 … –200
–200 … 0
0 … 1200

(8
0.06 °C + 0.05 % RDG
0.06 °C + 0.003 % RDG

(4
0.08 °C + 0.06 % RDG
0.08 °C + 0.006 % RDG

K (3

–270 … 1372

–270 … –200
–200 … 0
0 … 1000
1000 … 1372

(8
0.08 °C + 0.07 % RDG
0.08 °C + 0.004 % RDG
0.012 % RDG

(4
0.1 °C + 0.1 % RDG
0.1 °C + 0.007 % RDG
0.017 % RDG

N (3

–270 … 1300

–270 … –200
–200 …–100
–100 … 0
0 … 800
800 … 1300

(8
0.15 % RDG
0.11 °C + 0.04 % RDG
0.11 °C
0.06 °C + 0.006 % RDG

(4
0.2 % RDG
0.15 °C + 0.05 % RDG
0.15 °C
0.07 °C + 0.01 % RDG

T (3

–270 … 400

–270 … –200
–200 … 0
0 … 400

(8
0.07 °C + 0.07 % RDG
0.07 °C

(4
0.1 °C + 0.1 % RDG
0.1 °C

U (5

–200 … 600

–200 … 0
0 … 600

0.07 °C + 0.05 % RDG
0.07 °C

0.1 °C + 0.07 % RDG
0.1 °C

L (5

–200 … 900

–200 … 0
0 … 900

0.06 °C + 0.025 % RDG
0.06 °C + 0.002 % RDG

0.08 °C + 0.04 % RDG
0.08 °C + 0.005 % RDG

C (6

0 … 2315

0 … 1000
1000 … 2315

0.22 °C
0.018 % RDG

0.3 °C
0.027 % RDG

G (7

0 … 2315

0 … 60
60 … 200
200 … 400
400 … 1500
1500 … 2315

(8
0.9 °C
0.4 °C
0.2 °C
0.014 % RDG

(4
1.0 °C
0.5 °C
0.3 °C
0.02 % RDG

D (6

0 … 2315

0 … 140
140 … 1200
1200 … 2100
2100 … 2315

0.3 °C
0.2 °C
0.016 % RDG
0.45 °C

0.4 °C
0.3 °C
0.024 % RDG
0.65 °C

Footnotes

Resolution 0.01 °C.
With internal reference junction please see separate specification.
Also other thermocouple types available as option, please contact Beamex.
1) Accuracy includes hysteresis, nonlinearity and repeatability (k=2).
2) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned period (k=2).
3) IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
4) ±0.007 % of thermovoltage + 4 μV
5) DIN 43710
6) ASTM E 988 – 96
7) ASTM E 1751 – 95e1
8) ±0.004 % of thermovoltage + 3 μV

Measurement input impedance

> 10 MΩ

Simulation maximum load current

5 mA

Simulation load effect

< 5 μV/mA

Supported units

°C, °F, Kelvin, °Ré, °Ra

Connector

TC1: Universal TC connector , TC2: TC Miniplug

RTD MEASUREMENT & SIMULATION

R1 & R2 measurement

SENSOR TYPE

RANGE (°C)

RANGE (°C)

ACCURACY (1

1 YEAR UNCERTAINTY (2

Pt50(385)

−200 … 850

–200 … 270
270 … 850

0.025 °C
0.009 % RDG

0.03 °C
0.012 % RDG

Pt100(375)
Pt100(385)
Pt100(389)
Pt100(391)
Pt100(3926)

–200 … 850

–200 … 0
0 … 850

0.011 °C
0.011 °C + 0.009 % RDG

0.015 °C
0.015 °C + 0.012 % RDG

Pt100(3923)

–200 … 600

–200 … 0
0 … 600

0.011 °C
0.011 °C + 0.009 % RDG

0.015 °C
0.015 °C + 0.012 % RDG

Pt200(385)

–200 … 850

–200 … –80
–80 … 0
0 … 260
260 … 850

0.007 °C
0.016 °C
0.016 °C + 0.009 % RDG
0.03 °C + 0.011 % RDG

0.01 °C
0.02 °C
0.02 °C + 0.012 % RDG
0.045 °C + 0.02 % RDG

Pt400(385)

–200 … 850

–200 … –100
–100 … 0
0 … 850

0.007 °C
0.015 °C
0.026 °C + 0.01 % RDG

0.01 °C
0.02 °C
0.045 °C + 0.019 % RDG

Pt500(385)

–200 … 850

–200 … –120
–120 … –50
–50 … 0
0 … 850

0.008 °C
0.013 °C
0.025 °C
0.025 °C + 0.01 % RDG

0.01 °C
0.02 °C
0.045 °C
0.045 °C + 0.019 % RDG

Pt1000(385)

–200 … 850

−200 … −150
−150 … −50
−50 … 0
0 … 850

0.007 °C
0.018 °C
0.022 °C
0.022 °C + 0.01 % RDG

0.008 °C
0.03 °C
0.04 °C
0.04 °C + 0.019 % RDG

Ni100(618)

–60 … 180

−60 … 0
0 … 180

0.009 °C
0.009 °C + 0.005 % RDG

0.012 °C
0.012 °C + 0.006 % RDG

Ni120(672)

–80 … 260

−80 … 0
0 … 260

0.009 °C
0.009 °C + 0.005 % RDG

0.012 °C
0.012 °C + 0.006 % RDG

Cu10(427)

–200 … 260

–200 … 260

0.012 °C

0.16 °C

R1 simulation

SENSOR TYPE

RANGE (°C)

RANGE (°C)

ACCURACY (1

1 YEAR UNCERTAINTY (2

Pt50(385)

−200 … 850

–200 … 270
270 … 850

0.055 °C
0.035 °C + 0.008 % RDG

0.11 °C
0.11 °C + 0.015 % RDG

Pt100(375)
Pt100(385)
Pt100(389)
Pt100(391)
Pt100(3926)

–200 … 850

–200 … 0
0 … 850

0.025 °C
0.025 °C + 0.007 % RDG

0.05 °C
0.05 °C + 0.014 % RDG

Pt100(3923)

–200 … 600

–200 … 0
0 … 600

0.025 °C
0.025 °C + 0.007 % RDG

0.05 °C
0.05 °C + 0.014 % RDG

Pt200(385)

–200 … 850

–200 … –80
–80 … 0
0 … 260
260 … 850

0.012 °C
0.02 °C
0.02 °C + 0.006 % RDG
0.03 °C + 0.011 % RDG

0.025 °C
0.035 °C
0.04 °C + 0.011 % RDG
0.06 °C + 0.02 % RDG

Pt400(385)

–200 … 850

–200 … –100
–100 … 0
0 … 850

0.01 °C
0.015 °C
0.027 °C + 0.01 % RDG

0.015 °C
0.03 °C
0.05 °C + 0.019 % RDG

Pt500(385)

–200 … 850

–200 … –120
–120 … –50
–50 … 0
0 … 850

0.008 °C
0.012 °C
0.026 °C
0.026 °C + 0.01 % RDG

0.015 °C
0.025 °C
0.05 °C
0.05 °C + 0.019 % RDG

Pt1000(385)

–200 … 850

−200 … −150
−150 … −50
−50 … 0
0 … 850

0.006 °C
0.017 °C
0.023 °C
0.023 °C + 0.01 % RDG

0.011 °C
0.03 °C
0.043 °C
0.043 °C + 0.019 % RDG

Ni100(618)

–60 … 180

−60 … 0
0 … 180

0.021 °C
0.019 °C

0.042 °C
0.037 °C + 0.001 % RDG

Ni120(672)

–80 … 260

−80 … 0
0 … 260

0.021 °C
0.019 °C

0.042 °C
0.037 °C + 0.001 % RDG

Cu10(427)

–200 … 260

–200 … 260

0.26 °C

0.52 °C

For platinum sensors ITS-90 and Callendar van Dusen coefficients can be programmed. Also other RTD types available as option, please contact Beamex.

RTD Measurement current

Pulsed, bi-directional 1 mA (0…500 Ω), 0.2 mA (> 500 Ω)

4-wire connection

Measurement specifications valid

3-wire measurement

Add 10 mΩ

Max resistance excitation current

5 mA (0 … 650 Ω). Iexc × Rsim < 3.25 V (650 … 4000 Ω)

Min resistance excitation current

> 0.2 mA (0 … 400 Ω). > 0.1 mA (400 … 4000 Ω)

Simulation settling time with pulsed excitation current

< 1 ms

Supported units

°C, °F, Kelvin, °Ré, °Ra

INTERNAL REFERENCE JUNCTION TC1 & TC2

RANGE (°C)

ACCURACY (1

1 YEAR UNCERTAINTY (2

–10 … 45 °C

±0.10 °C

±0.15 °C

Specifications valid in temperature range: 15 … 35 °C.
Temperature coefficient outside of 15 … 35 °C: ±0.005 °C/ °C.
Specifications assumes that calibrator has stabilized in environmental condition, being switched on, for minimum of 90 minutes. For a measurement or simulation done sooner than that, please add uncertainty of 0.15 °C.
In order to calculate the total uncertainty of thermocouple measurement or simulation with internal reference junction used, please add the relevant thermocouple uncertainty and the reference junction uncertainty together as a root sum of the squares.

VOLTAGE MEASUREMENT

IN (–1 … 60 V)

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

–1.01 … 1 V

0.001 mV

3 μV + 0.003 % RDG

5 μV + 0.006 % RDG

1 … 60.6 V

0.01 mV

0.125 mV + 0.003 % RDG

0.25 mV + 0.006 % RDG

Input impedance

2 MΩ

Supported units

V, mV, μV

TC1 & TC2 (–1 … 1 V)

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

–1.01 … 1.01 V

0.001 mV

3 μV + 0.004 % RDG

4 μV + 0.007 % RDG

Input impedance

>10 MΩ

Supported units

V, mV, μV

Connector

TC1: Universal TC connector , TC2: TC Miniplug

VOLTAGE GENERATION

OUT (–3 … 24 V)

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

–3 … 10 V

0.00001 V

0.05 mV + 0.004 % RDG

0.1 mV + 0.007 % RDG

10 … 24 V

0.0001 V

0.05 mV + 0.004 % RDG

0.1 mV + 0.007 % RDG

Maximum load current

10 mA

Short circuit current

> 100 mA

Load effect

< 50 μV/mA

Supported units

V, mV, μV

TC1 (–1 … 1 V)

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

–1 … 1 V

0.001 mV

3 μV + 0.004 % RDG

4 μV + 0.007 % RDG

Maximum load current

5 mA

Load effect

< 5 μV/mA

Supported units

V, mV, μV

CURRENT MEASUREMENT

IN (–100 … 100 mA)

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

−25 … 25 mA

0.0001 mA

0.75 μA + 0.0075 % RDG

1 μA + 0.01 % RDG

±(25 … 101 mA)

0.001 mA

0.75 μA + 0.0075 % RDG

1 μA + 0.01 % RDG

Input impedance

< 10 Ω

Supported units

< 5 μV/mA

Supported units

mA, μA

Loop supply

Internal 24 V ±10 % (max 55 mA), or external max 60 VDC

CURRENT GENERATION

OUT (0 … 55 mA)

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

0 … 25 mA

0.0001 mA

0.75 μA + 0.0075 % RDG

1 μA + 0.01 % RDG

25 … 55 mA

0.001 mA

1.5 μA + 0.0075 % RDG

2 μA + 0.01 % RDG

Internal loop supply

24 V ±5 %. Max 55 mA

Max load impedance w. internal supply

24 V / (generated current). 1140 Ω @ 20 mA, 450 Ω @ 50 mA

Max external loop supply

60 VDC

Supported units

mA, μA

FREQUENCY MEASUREMENT

IN (0.0027 … 51 000 Hz)

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

0.0027 … 0.5 Hz

0.000001 Hz

0.000002 Hz + 0.001 % RDG

0.000002 Hz + 0.002 % RDG

0.5 … 5 Hz

0.00001 Hz

0.00002 Hz + 0.001 % RDG

0.00002 Hz + 0.002 % RDG

5 … 50 Hz

0.0001 Hz

0.0002 Hz + 0.001 % RDG

0.0002 Hz + 0.002 % RDG

50 … 500 Hz

0.001 Hz

0.002 Hz + 0.001 % RDG

0.002 Hz + 0.002 % RDG

500 … 5 000 Hz

0.01 Hz

0.02 Hz + 0.001 % RDG

0.02 Hz + 0.002 % RDG

5 000 … 51 000 Hz

0.1 Hz

0.2 Hz + 0.001 % RDG

0.2 Hz + 0.002 % RDG

Input impedance

> 1 MΩ

Supported units

Hz, kHz, cph, cpm, 1/Hz(s), 1/kHz(ms), 1/MHz(μs)

Trigger level

Dry contact, wet contact –1 … 14 V

Minimum signal amplitude

1.0 Vpp (<10kHz), 1.2 Vpp (10 … 50 kHz)

FREQUENCY GENERATION

OUT (0.0005 … 50 000 Hz)

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

0.0005 … 0.5 Hz

0.000001 Hz

0.000002 Hz + 0.001 % RDG

0.000002 Hz + 0.002 % RDG

0.5 … 5 Hz

0.00001 Hz

0.00002 Hz + 0.001 % RDG

0.00002 Hz + 0.002 % RDG

5 … 50 Hz

0.0001 Hz

0.0002 Hz + 0.001 % RDG

0.0002 Hz + 0.002 % RDG

50 … 500 Hz

0.001 Hz

0.002 Hz + 0.001 % RDG

0.002 Hz + 0.002 % RDG

500 … 5 000 Hz

0.01 Hz

0.02 Hz + 0.001 % RDG

0.02 Hz + 0.002 % RDG

5 000 … 50 000 Hz

0.1 Hz

0.2 Hz + 0.001 % RDG

0.2 Hz + 0.002 % RDG

Maximum load current

10 mA

Vawe forms

Positive square, symmetric square

Output amplitude positive square wave

0 … 24 Vpp

Output amplitude symmetric square wave

0 … 6 Vpp

Duty Cycle

1 … 99 %

Amplitude accuracy

< 5 % of amplitude

Supported units

Hz, kHz, cph, cpm, 1/Hz(s), 1/kHz(ms), 1/MHz(μs)

PULSE COUNTING

IN (0 … 9 999 999 pulses)

FEATURE

SPECIFICATION

Input impedance

> 1 MΩ

Trigger level

Dry contact, wet contact –1 … 14 V

Minimum signal amplitude

1 Vpp (< 10 kHz), 1.2 Vpp (10 … 50 kHz)

Max frequency

50 kHz

Trigger edge

Rising, falling

PULSE GENERATION

OUT (0 … 9 999 999 pulses)

FEATURE

SPECIFICATION

Resolution

1 pulse

Maximum load current

10 mA

Output amplitude positive pulse

0 … 24 Vpp

Output amplitude symmetric pulse

0 … 6 Vpp

Pulse frequency range

0.0005 … 10 000 Hz

Duty cycle

1 … 99 %

RESISTANCE MEASUREMENT

R1 & R2 (0 … 4000 Ω)

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

–1 … 100 Ω

0.001 Ω

4.5 mΩ

6 mΩ

100 … 110 Ω

0.001 Ω

0.0045 % RDG

0.006 % RDG

110 … 150 Ω

0.001 Ω

0.005 % RDG

0.007 % RDG

150 … 300 Ω

0.001 Ω

0.006 % RDG

0.008 % RDG

300 … 400 Ω

0.001 Ω

0.007 % RDG

0.009 % RDG

400 … 4 040 Ω

0.01 Ω

9 mΩ + 0.008 % RDG

12 mΩ + 0.015 % RDG

Measurement current

Pulsed, bi-directional 1 mA (0…500 Ω), 0.2 mA (>500 Ω)

Supported units

Positive square, symmetric square

4-wire connection

Measurement specifications valid

3-wire measurement

Add 10 mΩ

RESISTANCE SIMULATION

R1 (0 … 4000 Ω)

RANGE

RESOLUTION

ACCURACY (1

1 YEAR UNCERTAINTY (2

0 … 100 Ω

0.001 Ω

10 mΩ

20 mΩ

100 … 400 Ω

0.001 Ω

5 mΩ + 0.005 % RDG

10 mΩ + 0.01 % RDG

400 … 4 000 Ω

0.01 Ω

10 mΩ + 0.008 % RDG

20 mΩ + 0.015 % RDG

Max resistance excitation current

5 mA (0 … 650 Ω). Iexc × Rsim < 3.25 V (650 … 4000 Ω)

Min resistance excitation current

> 0.2 mA (0 … 400 Ω). >0.1 mA (400 … 4000 Ω)

Settling time with pulsed exitation current

< 1ms

Supported units

Ω, kΩ

1) Accuracy includes hysteresis, nonlinearity and repeatability (k=2).
2) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned period (k=2).