GENERAL SPECIFICATIONS
|
FEATURE |
VALUE |
|---|---|
|
Display |
5.7” Diagonal 640 x 480 TFT LCD module |
|
Touch panel |
5-wire resistive touch screen |
|
Keyboard |
Membrane keyboard |
|
Backlight |
LED backlight, adjustable brightness |
|
Weight |
Extended case: 1.5…2.0 kg (3.3…4.4 lb) Flat case: 1.5 kg (3.3 lb) |
|
Dimensions |
Extended case: 200 mm × 230 mm × 70 mm (D × W × H) ( 7.87 in × 9.06 in × 2.76 in) |
|
Battery type |
Rechargeable lithium-ion polymer, 4200 mAh, 11.1V |
|
Charging time |
Approximately 4 hours |
|
Charger supply |
100…240 VAC, 50–60 Hz |
|
Battery operation |
10…16 hours |
|
Operating temperature |
−10…45 °C (14…113 °F) |
|
Operating temperature while charging batteries |
0…30 °C (32…86 °F) |
|
Storage temperature |
−20…60 °C (−4…140 °F) |
|
Specifications valid |
−10…45 °C, unless other mentioned |
|
Humidity |
0…80% R.H. non condensing |
|
Warmup time |
Specifications valid after a 5 minute warmup period. |
|
Max. input voltage |
30 V AC, 60 V DC |
|
Display update rate |
3 readings/second |
|
Safety |
Directive 2014/35/EU, EN 61010-1:2010 + A1:2019 + AC:2019-04 |
|
EMC |
Directive 2014/30/EU, EN IEC 61326-1:2021 |
|
Ingress protection |
IP65 |
|
RoHS compliance |
ROHS II Directive 2011/65/EU, EN IEC 63000:2018 |
|
Drop |
IEC 60068-2-32. 1 meter (3.28 ft) |
|
Vibration |
IEC 60068-2-64. Random, 2 g, 5…500 Hz |
|
Max altitude |
3,000 m (9,842 ft) |
|
Warranty |
Warranty 3 years. 1 year for battery pack. |
PRESSURE MEASUREMENT
|
INTERNAL MODULES |
EXTERNAL MODULES |
UNIT |
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|---|---|---|
|
PB |
EXT B |
kPa a |
70 … 120 |
0.01 |
0.03 kPa |
0.05 kPa |
|
|
|
|
|
|
|
|
|
P10mD |
EXT10mD |
kPa mbar iwc |
±1 ±10 ±4 |
0.0001 |
0.05 % Span |
0.05 % Span + 0.1 % RDG |
|
P100m |
EXT100m |
kPa mbar iwc |
0 … 10 0 … 100 0 … 40 |
0.0001 |
0.015 % FS + 0.0125 % RDG |
0.025 % FS + 0.025 % RDG |
|
P400mC |
EXT400mC |
kPa mbar iwc |
±40 |
0.001 |
0.01 % FS + 0.0125 % RDG |
0.02 % FS + 0.025 % RDG |
|
P1C |
EXT1C |
kPa bar psi c |
±100 |
0.001 |
0.007 % FS + 0.0125 % RDG |
0.015 % FS + 0.025 % RDG |
|
P2C |
EXT2C |
kPa bar psi |
–100 … 200 |
0.001 |
0.005 % FS + 0.01 % RDG |
0.01 % FS + 0.025 % RDG |
|
P6C |
EXT6C |
kPa bar psi |
–100 … 600 |
0.01 |
0.005 % FS + 0.01 % RDG |
0.01 % FS + 0.025 % RDG |
|
P20C |
EXT20C |
kPa bar psi |
–100 … 2000 |
0.01 |
0.005 % FS + 0.01 % RDG |
0.01 % FS + 0.025 % RDG |
|
P60 |
EXT60 |
kPa bar psi |
0 … 6000 |
0.1 |
0.005 % FS + 0.0125 % RDG |
0.01 % FS + 0.025 % RDG |
|
P100 |
EXT100 |
MPa bar psi |
0 … 10 |
0.0001 |
0.005 % FS + 0.0125 % RDG |
0.01 % FS + 0.025 % RDG |
|
P160 |
EXT160 |
MPa bar psi |
0 … 16 |
0.0001 |
0.005 % FS + 0.0125 % RDG |
0.01 % FS + 0.025 % RDG |
|
– |
EXT250 |
MPa bar psi |
0 … 25 |
0.001 |
0.007 % FS + 0.0125 % RDG |
0.015 % FS + 0.025 % RDG |
|
– |
EXT600 |
MPa bar psi |
0 … 60 |
0.001 |
0.007 % FS + 0.01 % RDG |
0.015 % FS + 0.025 % RDG |
|
– |
EXT1000 |
MPa bar psi |
0 … 100 |
0.001 |
0.007 % FS + 0.01 % RDG |
0.015 % FS + 0.025 % RDG |
Footnotes
1) Accuracy includes hysteresis, nonlinearity and repeatability (k=2).
2) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned period (k=2).
3) Every internal/external gauge pressure module’s range may be displayed also in absolute pressure if the barometric module (PB or EXT B) is installed/connected.
Maximum number of internal pressure modules is 3 gauge/differential pressure modules and one barometric (PB) module.
There is a connection for external pressure modules.
% RDG = Percentage of the actual pressure reading.
% FS = Percentage of the full scale range from zero to maximum.
% Span = Percentage of the total measurement range from minimum to maximum.
SUPPORTED PRESSURE UNITS
Pa, kPa, hPa, MPa, mbar, bar, gf/cm2, kgf/cm2, kgf/m2, kp/cm2, lbf/ft2, psi,at, torr, atm, ozf/in2, iwc, inH2O, ftH2O, mmH2O, cmH2O, mH2O, mmHg, cmHg, mHg, inHg, mmHg(0 °C), inHg(0 °C), mmH2O(60°F), mmH2O(68°F), mmH2O(4 °C), cmH2O(60°F), cmH2O(68°F), cmH2O(4 °C), inH2O(60°F), inH2O(68°F), inH2O(4 °C), ftH2O(60°F), ftH2O(68°F), ftH2O(4 °C).
Large number of user pressure units can be created.
TEMPERATURE COEFFICIENT
<±0.001 % RDG / °C outside 15–35 °C (59–95 °F).
P10mD / EXT10mD: < ±0.002 % Span / °C outside 15–35 °C (59–95 °F).
MAX OVERPRESSURE
2 times the nominal pressure. Except the following modules;
PB/EXTB: 1200 mbar abs (35.4 inHg abs).
P10mD/EXT10mD: 200 mbar (80 iwc).
EXT600: 900 bar (13000 psi).
EXT1000: 1000 bar (15000 psi).
PRESSURE MEDIA
Modules up to P6C/EXT6C: dry, clean air or other clean, inert, non-toxic, noncorrosive gases.
Modules P20C/EXT20C and higher: clean, inert, non-toxic, non-corrosive gases or liquids.
WETTED PARTS
AISI316 stainless steel, Hastelloy, Nitrile rubber.
PRESSURE CONNECTIONS
PB/EXTB: 10/32″ (M5) female.
P10mD/EXT10mD: 2 x adapters for 1/8” ID hose (3,2 mm).
P100m/EXT100m to P20C/EXT20C: Bx G1/8” male compatible with Beamex 40 bar hoses.
P60, P100, P160: Bx 1215 male compatible with Beamex 630 bar hoses.
EXT60 to EXT1000: G1/4″ B male.
THERMOCOUPLE MEASUREMENT & SIMULATION
TC1 measurement & simulation / TC2 measurement
|
TYPE |
RANGE (°C) |
RANGE (°C) |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|---|
|
B (3 |
0 … 1820 |
0 … 200 |
(8 |
(4 |
|
R (3 |
–50 … 1768 |
–50 … 0 |
0.8 °C |
1.0 °C |
|
S (3 |
–50 … 1768 |
–50 … 0 |
0.7 °C |
0.9 °C |
|
E (3 |
–270 … 1000 |
–270 … –200 |
(8 |
(4 |
|
J (3 |
–210 … 1200 |
–210 … –200 |
(8 |
(4 |
|
K (3 |
–270 … 1372 |
–270 … –200 |
(8 |
(4 |
|
N (3 |
–270 … 1300 |
–270 … –200 |
(8 |
(4 |
|
T (3 |
–270 … 400 |
–270 … –200 |
(8 |
(4 |
|
U (5 |
–200 … 600 |
–200 … 0 |
0.07 °C + 0.05 % RDG |
0.1 °C + 0.07 % RDG |
|
L (5 |
–200 … 900 |
–200 … 0 |
0.06 °C + 0.025 % RDG |
0.08 °C + 0.04 % RDG |
|
C (6 |
0 … 2315 |
0 … 1000 |
0.22 °C |
0.3 °C |
|
G (7 |
0 … 2315 |
0 … 60 |
(8 |
(4 |
|
D (6 |
0 … 2315 |
0 … 140 |
0.3 °C |
0.4 °C |
Footnotes
Resolution 0.01 °C.
With internal reference junction please see separate specification.
Also other thermocouple types available as option, please contact Beamex.
1) Accuracy includes hysteresis, nonlinearity and repeatability (k=2).
2) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned period (k=2).
3) IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
4) ±0.007 % of thermovoltage + 4 μV
5) DIN 43710
6) ASTM E 988 – 96
7) ASTM E 1751 – 95e1
8) ±0.004 % of thermovoltage + 3 μV
|
Measurement input impedance |
> 10 MΩ |
|
Simulation maximum load current |
5 mA |
|
Simulation load effect |
< 5 μV/mA |
|
Supported units |
°C, °F, Kelvin, °Ré, °Ra |
|
Connector |
TC1: Universal TC connector , TC2: TC Miniplug |
RTD MEASUREMENT & SIMULATION
R1 & R2 measurement
|
SENSOR TYPE |
RANGE (°C) |
RANGE (°C) |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|---|
|
Pt50(385) |
−200 … 850 |
–200 … 270 |
0.025 °C |
0.03 °C |
|
Pt100(375) |
–200 … 850 |
–200 … 0 |
0.011 °C |
0.015 °C |
|
Pt100(3923) |
–200 … 600 |
–200 … 0 |
0.011 °C |
0.015 °C |
|
Pt200(385) |
–200 … 850 |
–200 … –80 |
0.007 °C |
0.01 °C |
|
Pt400(385) |
–200 … 850 |
–200 … –100 |
0.007 °C |
0.01 °C |
|
Pt500(385) |
–200 … 850 |
–200 … –120 |
0.008 °C |
0.01 °C |
|
Pt1000(385) |
–200 … 850 |
−200 … −150 |
0.007 °C |
0.008 °C |
|
Ni100(618) |
–60 … 180 |
−60 … 0 |
0.009 °C |
0.012 °C |
|
Ni120(672) |
–80 … 260 |
−80 … 0 |
0.009 °C |
0.012 °C |
|
Cu10(427) |
–200 … 260 |
–200 … 260 |
0.012 °C |
0.16 °C |
R1 simulation
|
SENSOR TYPE |
RANGE (°C) |
RANGE (°C) |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|---|
|
Pt50(385) |
−200 … 850 |
–200 … 270 |
0.055 °C |
0.11 °C |
|
Pt100(375) |
–200 … 850 |
–200 … 0 |
0.025 °C |
0.05 °C |
|
Pt100(3923) |
–200 … 600 |
–200 … 0 |
0.025 °C |
0.05 °C |
|
Pt200(385) |
–200 … 850 |
–200 … –80 |
0.012 °C |
0.025 °C |
|
Pt400(385) |
–200 … 850 |
–200 … –100 |
0.01 °C |
0.015 °C |
|
Pt500(385) |
–200 … 850 |
–200 … –120 |
0.008 °C |
0.015 °C |
|
Pt1000(385) |
–200 … 850 |
−200 … −150 |
0.006 °C |
0.011 °C |
|
Ni100(618) |
–60 … 180 |
−60 … 0 |
0.021 °C |
0.042 °C |
|
Ni120(672) |
–80 … 260 |
−80 … 0 |
0.021 °C |
0.042 °C |
|
Cu10(427) |
–200 … 260 |
–200 … 260 |
0.26 °C |
0.52 °C |
For platinum sensors ITS-90 and Callendar van Dusen coefficients can be programmed. Also other RTD types available as option, please contact Beamex.
|
RTD Measurement current |
Pulsed, bi-directional 1 mA (0…500 Ω), 0.2 mA (> 500 Ω) |
|
4-wire connection |
Measurement specifications valid |
|
3-wire measurement |
Add 10 mΩ |
|
Max resistance excitation current |
5 mA (0 … 650 Ω). Iexc × Rsim < 3.25 V (650 … 4000 Ω) |
|
Min resistance excitation current |
> 0.2 mA (0 … 400 Ω). > 0.1 mA (400 … 4000 Ω) |
|
Simulation settling time with pulsed excitation current |
< 1 ms |
|
Supported units |
°C, °F, Kelvin, °Ré, °Ra |
INTERNAL REFERENCE JUNCTION TC1 & TC2
|
RANGE (°C) |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|
|
–10 … 45 °C |
±0.10 °C |
±0.15 °C |
Specifications valid in temperature range: 15 … 35 °C.
Temperature coefficient outside of 15 … 35 °C: ±0.005 °C/ °C.
Specifications assumes that calibrator has stabilized in environmental condition, being switched on, for minimum of 90 minutes. For a measurement or simulation done sooner than that, please add uncertainty of 0.15 °C.
In order to calculate the total uncertainty of thermocouple measurement or simulation with internal reference junction used, please add the relevant thermocouple uncertainty and the reference junction uncertainty together as a root sum of the squares.
VOLTAGE MEASUREMENT
IN (–1 … 60 V)
|
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|
|
–1.01 … 1 V |
0.001 mV |
3 μV + 0.003 % RDG |
5 μV + 0.006 % RDG |
|
1 … 60.6 V |
0.01 mV |
0.125 mV + 0.003 % RDG |
0.25 mV + 0.006 % RDG |
|
Input impedance |
2 MΩ |
|
Supported units |
V, mV, μV |
TC1 & TC2 (–1 … 1 V)
|
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|
|
–1.01 … 1.01 V |
0.001 mV |
3 μV + 0.004 % RDG |
4 μV + 0.007 % RDG |
|
Input impedance |
>10 MΩ |
|
Supported units |
V, mV, μV |
|
Connector |
TC1: Universal TC connector , TC2: TC Miniplug |
VOLTAGE GENERATION
OUT (–3 … 24 V)
|
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|
|
–3 … 10 V |
0.00001 V |
0.05 mV + 0.004 % RDG |
0.1 mV + 0.007 % RDG |
|
10 … 24 V |
0.0001 V |
0.05 mV + 0.004 % RDG |
0.1 mV + 0.007 % RDG |
|
Maximum load current |
10 mA |
|
Short circuit current |
> 100 mA |
|
Load effect |
< 50 μV/mA |
|
Supported units |
V, mV, μV |
TC1 (–1 … 1 V)
|
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|
|
–1 … 1 V |
0.001 mV |
3 μV + 0.004 % RDG |
4 μV + 0.007 % RDG |
|
Maximum load current |
5 mA |
|
Load effect |
< 5 μV/mA |
|
Supported units |
V, mV, μV |
CURRENT MEASUREMENT
IN (–100 … 100 mA)
|
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|
|
−25 … 25 mA |
0.0001 mA |
0.75 μA + 0.0075 % RDG |
1 μA + 0.01 % RDG |
|
±(25 … 101 mA) |
0.001 mA |
0.75 μA + 0.0075 % RDG |
1 μA + 0.01 % RDG |
|
Input impedance |
< 10 Ω |
|
Supported units |
< 5 μV/mA |
|
Supported units |
mA, μA |
|
Loop supply |
Internal 24 V ±10 % (max 55 mA), or external max 60 VDC |
CURRENT GENERATION
OUT (0 … 55 mA)
|
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|
|
0 … 25 mA |
0.0001 mA |
0.75 μA + 0.0075 % RDG |
1 μA + 0.01 % RDG |
|
25 … 55 mA |
0.001 mA |
1.5 μA + 0.0075 % RDG |
2 μA + 0.01 % RDG |
|
Internal loop supply |
24 V ±5 %. Max 55 mA |
|
Max load impedance w. internal supply |
24 V / (generated current). 1140 Ω @ 20 mA, 450 Ω @ 50 mA |
|
Max external loop supply |
60 VDC |
|
Supported units |
mA, μA |
FREQUENCY MEASUREMENT
IN (0.0027 … 51 000 Hz)
|
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|
|
0.0027 … 0.5 Hz |
0.000001 Hz |
0.000002 Hz + 0.001 % RDG |
0.000002 Hz + 0.002 % RDG |
|
0.5 … 5 Hz |
0.00001 Hz |
0.00002 Hz + 0.001 % RDG |
0.00002 Hz + 0.002 % RDG |
|
5 … 50 Hz |
0.0001 Hz |
0.0002 Hz + 0.001 % RDG |
0.0002 Hz + 0.002 % RDG |
|
50 … 500 Hz |
0.001 Hz |
0.002 Hz + 0.001 % RDG |
0.002 Hz + 0.002 % RDG |
|
500 … 5 000 Hz |
0.01 Hz |
0.02 Hz + 0.001 % RDG |
0.02 Hz + 0.002 % RDG |
|
5 000 … 51 000 Hz |
0.1 Hz |
0.2 Hz + 0.001 % RDG |
0.2 Hz + 0.002 % RDG |
|
Input impedance |
> 1 MΩ |
|
Supported units |
Hz, kHz, cph, cpm, 1/Hz(s), 1/kHz(ms), 1/MHz(μs) |
|
Trigger level |
Dry contact, wet contact –1 … 14 V |
|
Minimum signal amplitude |
1.0 Vpp (<10kHz), 1.2 Vpp (10 … 50 kHz) |
FREQUENCY GENERATION
OUT (0.0005 … 50 000 Hz)
|
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|
|
0.0005 … 0.5 Hz |
0.000001 Hz |
0.000002 Hz + 0.001 % RDG |
0.000002 Hz + 0.002 % RDG |
|
0.5 … 5 Hz |
0.00001 Hz |
0.00002 Hz + 0.001 % RDG |
0.00002 Hz + 0.002 % RDG |
|
5 … 50 Hz |
0.0001 Hz |
0.0002 Hz + 0.001 % RDG |
0.0002 Hz + 0.002 % RDG |
|
50 … 500 Hz |
0.001 Hz |
0.002 Hz + 0.001 % RDG |
0.002 Hz + 0.002 % RDG |
|
500 … 5 000 Hz |
0.01 Hz |
0.02 Hz + 0.001 % RDG |
0.02 Hz + 0.002 % RDG |
|
5 000 … 50 000 Hz |
0.1 Hz |
0.2 Hz + 0.001 % RDG |
0.2 Hz + 0.002 % RDG |
|
Maximum load current |
10 mA |
|
Vawe forms |
Positive square, symmetric square |
|
Output amplitude positive square wave |
0 … 24 Vpp |
|
Output amplitude symmetric square wave |
0 … 6 Vpp |
|
Duty Cycle |
1 … 99 % |
|
Amplitude accuracy |
< 5 % of amplitude |
|
Supported units |
Hz, kHz, cph, cpm, 1/Hz(s), 1/kHz(ms), 1/MHz(μs) |
PULSE COUNTING
IN (0 … 9 999 999 pulses)
|
FEATURE |
SPECIFICATION |
|---|---|
|
Input impedance |
> 1 MΩ |
|
Trigger level |
Dry contact, wet contact –1 … 14 V |
|
Minimum signal amplitude |
1 Vpp (< 10 kHz), 1.2 Vpp (10 … 50 kHz) |
|
Max frequency |
50 kHz |
|
Trigger edge |
Rising, falling |
PULSE GENERATION
OUT (0 … 9 999 999 pulses)
|
FEATURE |
SPECIFICATION |
|---|---|
|
Resolution |
1 pulse |
|
Maximum load current |
10 mA |
|
Output amplitude positive pulse |
0 … 24 Vpp |
|
Output amplitude symmetric pulse |
0 … 6 Vpp |
|
Pulse frequency range |
0.0005 … 10 000 Hz |
|
Duty cycle |
1 … 99 % |
RESISTANCE MEASUREMENT
R1 & R2 (0 … 4000 Ω)
|
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|
|
–1 … 100 Ω |
0.001 Ω |
4.5 mΩ |
6 mΩ |
|
100 … 110 Ω |
0.001 Ω |
0.0045 % RDG |
0.006 % RDG |
|
110 … 150 Ω |
0.001 Ω |
0.005 % RDG |
0.007 % RDG |
|
150 … 300 Ω |
0.001 Ω |
0.006 % RDG |
0.008 % RDG |
|
300 … 400 Ω |
0.001 Ω |
0.007 % RDG |
0.009 % RDG |
|
400 … 4 040 Ω |
0.01 Ω |
9 mΩ + 0.008 % RDG |
12 mΩ + 0.015 % RDG |
|
Measurement current |
Pulsed, bi-directional 1 mA (0…500 Ω), 0.2 mA (>500 Ω) |
|
Supported units |
Positive square, symmetric square |
|
4-wire connection |
Measurement specifications valid |
|
3-wire measurement |
Add 10 mΩ |
RESISTANCE SIMULATION
R1 (0 … 4000 Ω)
|
RANGE |
RESOLUTION |
ACCURACY (1 |
1 YEAR UNCERTAINTY (2 |
|---|---|---|---|
|
0 … 100 Ω |
0.001 Ω |
10 mΩ |
20 mΩ |
|
100 … 400 Ω |
0.001 Ω |
5 mΩ + 0.005 % RDG |
10 mΩ + 0.01 % RDG |
|
400 … 4 000 Ω |
0.01 Ω |
10 mΩ + 0.008 % RDG |
20 mΩ + 0.015 % RDG |
|
Max resistance excitation current |
5 mA (0 … 650 Ω). Iexc × Rsim < 3.25 V (650 … 4000 Ω) |
|
Min resistance excitation current |
> 0.2 mA (0 … 400 Ω). >0.1 mA (400 … 4000 Ω) |
|
Settling time with pulsed exitation current |
< 1ms |
|
Supported units |
Ω, kΩ |
1) Accuracy includes hysteresis, nonlinearity and repeatability (k=2).
2) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned period (k=2).