INTERNAL MODULES EXTERNAL MODULES RANGE RESOLUTION ACCURACY 1 YEAR UNCERTAINTY
PB EXT B 10.15 to 17.4 psi a 0.0001 .03 mbar 0.0073 psi
P10mD EXT10mD ±4 iwc diff 0.001 0.05% span 0.05% span + 0.1 % RDG
P100m EXT100m 0 to 40 iwc 0.001 0.015% FS + 0.0125% RDG 0.025% FS + 0.025% RDG
P400mC EXT400mC ±160 iwc 0.001 0.01% FS+ 0.0125% RDG 0.02% FS + 0.025% RDG
P1C EXT1C – 14.5 to 15 psi 0.0001 0.007% FS  + 0.0125% RDG 0.015% FS + 0.025% RDG
P2C EXT2C – 14.5 to 30 psi 0.0001 0.005% FS + 0.01% RDG 0.01% FS +  0.025% RDG
P6C EXT6C – 14.5 to 90 psi 0.001 0.005% FS + 0.01% RDG 0.01% FS +  0.025% RDG
P20C EXT20C – 14.5 to 300 psi 0.001 0.005% FS + 0.01% RDG 0.01% FS +  0.025% RDG
P60 EXT60 0 to 900 psi 0.01 0.005% FS + 0.0125% RDG 0.01% FS +  0.025% RDG
P100 EXT100 0 to 1,500 psi 0.01 0.005% FS + 0.0125% RDG 0.01% FS +  0.025% RDG
P160 EXT160 0 to 2,400 psi 0.01 0.005% FS + 0.0125% RDG 0.01% FS +  0.025% RDG
EXT250 0 to 3,700 psi 0.1 0.007% FS + 0.0125% RDG 0.015% FS +  0.025% RDG
EXT600 0 to 9,000 psi 0.1 0.007% FS + 0.01% RDG 0.015% FS +  0.025% RDG
EXT1000 0 to 15,000 psi 0.1 0.007% FS + 0.01% RDG 0.015% FS +  0.025% RDG

 

1) Accuracy includes hysteresis, nonlinearity and repeatability (k=2).
2) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned period (k=2).
3) Every internal/external gauge pressure module’s range may be displayed also in absolute pressure if the barometric module (PB or EXT B) is installed/connected.
Maximum number of internal pressure modules is 3 gauge/differential pressure modules and one barometric (PB) module in the extended case. The flat case has room for internal barometric module only. Both cases have connection for external pressure modules.
External pressure modules are also compatible with Beamex MC2, MC4 and MC5 family calibrators.