IN (-100 … 100 mA)

RANGE RESOLUTION ACCURACY 1 YEAR UNCERTAINTY
−25 … 25 mA 0.0001 mA 0.75 µA + 0.0075 % RDG 1 µA + 0.01 % RDG
±(25 … 101 mA) 0.001 mA 0.75 µA + 0.0075 % RDG 0.75 µA + 0.0075 % RDG
Input impedance < 10 Ω
Supported units mA, µA
Loop Supply Internal 24 V ±10 % (max 55 mA), or external max 60 VDC

 

Accuracy includes hysteresis, nonlinearity and repeatability (k=2).

Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned period (k=2).