IN (–1…60 V)

RANGE RESOLUTION ACCURACY 1 YEAR UNCERTAINTY
– 1.01… 1 V 0.001mV 3 µV + 0.003% RDG 5 µV + 0.006% RDG
1…60.6 V 0.01 mV 0.125 mV + 0.003% RDG 0.25 mV + 0.006% RDG
Input impedance > 10 MΩ
Supported units V, mV, µV
Connector TC1: Universal TC connector , TC2: TC Miniplug

 

 

TC1 & TC2 (–1…1 V)

RANGE RESOLUTION ACCURACY 1 YEAR UNCERTAINTY
–1.01…1.01 V 0.001 mV 3 µV + 0.004% RDG 4 µV + 0.007% RDG
Input impedance > 10 MΩ
Supported units V, mV, µV
Connector TC1: Universal TC connector , TC2: TC Miniplug

 

1) Accuracy includes hysteresis, nonlinearity and repeatability (k=2).
2) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned period (k=2).