mV measurement 2) |
± 1000 mV |
0.001 - 0.01 mV |
0.02% RDG + 5 µV |
| |
V measurement
3) |
± 50 V |
0.00001 - 0.001
V |
0.02% RDG + 0.25
mV |
| |
mA measurement 4) |
± 100 mA |
0.0001 - 0.001 mA |
0.02% RDG + 1.5 µA |
| |
Hz measurement
5) |
0.0028 to 50000
Hz |
0.000001 - 0.1
Hz |
0.01% RDG |
| |
Pulse counting 5) |
0 to 9 999 999 pulses |
1 pulse |
N/A |
| |
mA generation
6) |
0 to 25 mA |
0.0001 mA |
0.02% RDG + 1.5
µA |
1) Uncertainty includes reference standard uncertainty,
hysteresis, nonlinearity,
repeatability and typical long term stability for mentioned period.(k=2)
2) Bias current <10 nA
3) Impedance >1 Mohm
4) Impedance <7.5 ohm
5) Impedance >1 Mohm. Frequency measurement minimum amplitude 0.5 Vpp
(<5 kHz),1Vpp (5...50 kHz).Pulse counting minimum amplitude 0.5 Vpp
(pulse length >100 µs),1 Vpp (pulse length 100 µs...10
µs).
Trigger level range -1
+15 V.
6) Maximum load impedance 800 ohm
| |
mV generation 2) |
± 500 mV |
0.001 - 0.01 mV |
0.02% RDG + 4 µV |
| |
V generation 3) |
± 12 V |
0.00001 - 0.0001 V |
0.02% RDG + 0.1 mV |
| |
mA generation 4) |
± 25 mA |
0.0001 mA |
0.02% RDG + 1 µA |
| |
Hz generation 5) |
0.00028 to 50000 Hz |
0.000001 - 0.1 Hz |
0.01% RDG |
| |
Pulse generation 6) |
0 to 9 999 999 pulses |
1 pulse |
N/A |
| |
Ohm simulation 7) |
1 to 4000 ohm |
0.01 - 0.1 ohm |
0.04% RDG or 30 mohm 8) |
| |
Ohm measurement 9) |
0 to 4000 ohm |
0.001 - 0.1 ohm |
0.02% RDG + 3.5 mohm |
| |
mV measurement 10) |
± 500 mV |
0.001 - 0.01 mV |
0.02% RDG + 4 µV |
1) Uncertainty includes reference standard uncertainty,
hysteresis, nonlinearity,
repeatability and typical long term stability for mentioned period.(k=2)
2) Load effect <5 µV/mA. Maximum output current 5 mA.
3) Load effect <100 µV/mA. Maximum output current 10 mA (0 ..10
V),
3 mA (10 ..12 V).
4) Maximum load impedance 400 ohm.
5) Amplitude range 0 ..12 Vpp. Amplitude setting accuracy up to 5 kHz
±(200 mV +5%of set value). Waveforms: Square wave
(positive /symmetric)and sinewave (above 40 Hz).
6) Pulse generation frequency range 0.1 ...1000 Hz.
Amplitude setting 0 ...12 Vpp.
7) Valid with measurement current 0.2 ...5 mA (1 ...1000 ohm),0.1 ...1mA
(1 ...4 kohm).Ohm/RTD simulation speed 1 ms.
8) Which ever is greater.
9) Specification valid with 4 wire connection.In 3 wire connection add
10 mohm.
10) Bias current <10 nA.
| |
Pt-sensors |
-200 to 850°C |
-200 to 0°C |
0.06°C
|
0.1°C |
| |
|
|
0 to 850°C |
0.025% RDG + 0.06°C |
0.025% RDG + 0.1°C |
| Pt50 (385) |
Pt500 (385) |
Pt100 (391) |
Ni100 (618) |
| Pt100 (385) |
Pt1000 (385) |
Pt100 (375) |
Ni120 (672) |
| Pt200 (385) |
Pt100 (3923) |
Pt100 (3926) |
Cu10 (427) |
| Pt400 (385) |
Pt100 (389) |
|
|
RTD/ohm simulation excitation current 0.2
5 mA (1
1000 ohm), 0.1
1 mA (1 ... 4 kohm).
Also other RTD types available as option.
| |
B 2) |
0 to 1820 |
0 to 200
200 to 500
500 to 800
800 to 1820 |
3)
2.0°C
0.6°C
0.8°C |
| |
R 2) |
-50 to 1768 |
-50 to 0
0 to 150
150 to 1400
1400 to 1768 |
1.0°C
0.7°C
0.5°C
0.6°C |
| |
S 2) |
-50 to 1768 |
-50 to 0
0 to 50
50 to 1500
1500 to 1768 |
1.0°C
0.7°C
0.6°C
0.7°C |
| |
E 2) |
-270 to 1000 |
-270 to -200
-200 to 0
0 to 600
600 to 1000 |
3)
0.08% RDG + 0.07°C
0.015% RDG + 0.07°C
0.026% RDG |
| |
J 2) |
-210 to 1200 |
-210 to -200
-200 to 0
0 to 1200 |
3)
0.07%RDG + 0.08°C
0.02% RDG + 0.08°C |
| |
K 2) |
-270 to 1372 |
-270 to -200
-200 to 0
0 to 1000
1000 to 1372 |
3)
0.1% RDG + 0.1°C
0.02% RDG + 0.1°C
0.03% RDG |
| |
N 2) |
-270 to 1300 |
-270 to -200
-200 to -100
-100 to 0
0 to 750
750 to 1300 |
3)
0.2% RDG
0.05% RDG + 0.15°C
0.01% RDG + 0.15°C
0.03% RDG |
| |
T 2) |
-270 to 400 |
-270 to -250
-250 to -200
-200 to 0
0 to 400 |
3)
0.7°C
0.1% RDG + 0.1°C
0.01% RDG + 0.1°C |
| |
U 4) |
200 to 600 |
-200 to 0
0 to 600 |
0.1% RDG + 0.15°C
0.01% RDG + 0.15°C |
| |
L 4) |
-200 to 900 |
-200 to 0
0 to 900 |
0.07% RDG + 0.13°C
0.02% RDG + 0.13°C |
Resolution 0.01°C.
With internal reference junction (module RJ) add 0.1°C uncertainty.
Thermocouple types C 3) (ASTM E 988 - 96), G 3)
(ASTM E 1751 - 95e1) and D 3) (ASTM E 988 - 96) also available as standard.
Also other thermocouple types available as option.
1) Uncertainty includes reference standard
uncertainty, hysteresis, nonlinearity, repeatability and typical long
term stability for
mentioned period. (k=2)
2) IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
3) ±(0.02 % RDG + 4) mV
4) DIN 43710
| |
PM B
EXT B
|
kPa a
mbar a
psi a |
80 to 120
800 to 1200
11.6 to 17.4 |
0.01
0.1
0.001 |
-
-
-
|
0.05 kPa
0.5 mbar
0.0073 psi |
| |
PM10mD
EXT10mD |
kPa
mba
riwc |
± 1
± 10
± 4 |
0.0001
0.001
0.001 |
-
-
- |
0.05% span |
| |
PM100m
EXT100m |
kPa
mbar
iwc |
0 to 10
0 to 100
0 to 40 |
0.0001
0.001
0.001 |
0.025% RDG + 0.025% FS |
0.04% RDG + 0.025% FS |
| |
PM400mC
EXT400mC |
kPa
mbar
psi |
± 40
± 400
± 160 |
0.001
0.01
0.001 |
0.025% RDG +
0.02% FS |
0.04% RDG + 0.02%
FS |
| |
PM1C
EXT1C |
kPa
bar
psi |
± 100
± 1
-14.5 to 15 |
0.001
0.00001
0.0001 |
0.025% RDG + 0.01% FS |
0.04% RDG + 0.01% FS |
| |
PM2C
EXT2C |
kPa
bar
psi |
-100 to 200
-1 to 2
-14.5 to 30 |
0.001
0.00001
0.0001 |
0.025% RDG + 0.01% FS |
0.04% RDG + 0.01% FS |
| |
PM6C
EXT6C |
kPa
bar
psi |
-100 to 600
-1 to 6
-14.5 to 90 |
0.01
0.0001
0.001 |
0.025% RDG + 0.01% FS |
0.04% RDG + 0.01% FS |
| |
PM20C
EXT20C |
kPa
bar
psi |
-100 to 2000
-1 to 20
-14.5 to 300 |
0.01
0.0001
0.001 |
0.025% RDG + 0.01% FS |
0.04% RDG + 0.01% FS |
| |
PM60
EXT60 |
kPa
bar
psi |
0 to 6000
0 to 60
0 to 900 |
0.1
0.001
0.01 |
-
-
-
|
0.04% RDG + 0.01% FS |
| |
PM100
EXT100 |
MPa
bar
psi |
0 to 10
0 to 100
0 to 1500 |
0.0001
0.001
0.01 |
-
-
- |
0.04% RDG + 0.01% FS |
| |
PM160
EXT160 |
MPa
bar
psi |
0 to 16
0 to 160
0 to 2400 |
0.0001
0.001
0.01 |
-
-
- |
0.04% RDG + 0.013% FS |
| |
PM250
EXT250 |
MPa
bar
psi |
0 to 25
0 to 250
0 to 3700 |
0.001
0.01
0.1 |
-
-
- |
0.04% RDG + 0.015% FS |
| |
PM600
EXT600 |
MPa
bar
psi |
0 to 60
0 to 600
0 to 9000 |
0.001
0.01
0.1 |
-
-
- |
0.04% RDG + 0.015% FS |
| |
-
EXT1000 |
MPa
bar
psi |
0 to 100
0 to 1000
0 to 15000 |
0.001
0.01
0.1 |
-
-
- |
0.04% RDG + 0.015% FS |
1) Uncertainty includes reference standard uncertainty,
hysteresis, nonlinearity, repeatability and typical long term stability
for mentioned period. (k=2)
2) Every internal/external pressure module's range may be displayed also
in absolute pressure if the Barometric Module (B) is installed.
Supports the following pressure units as standard:
Pa, hPa, kPa, MPa, mbar, bar, lbf/ft 2 , psi, gf/cm 2 , kgf/cm 2 , kgf/m
2 , kp/cm 2 , at, mmH 2 O, cmH 2 O, mH 2 O, iwc, ftH 2 O, mmHg, cmHg,
mHg, inHg, mmHg(0°C), inHg(0°C), mmH 2 O(4°C), inH 2 O(4°C),
ftH 2 O(4°C), inH 2 O(60°F), mmH 2 O(68°F), inH 2 O(68°F),
ftH 2 O(68°F), torr, atm.
Pressure modules 20bar and below, pressure connection
G 1/8 (ISO 228/1) 60° internal cone.
PM60, PM100, PM160, PM250, PM600 module pressure connection G 1/8 (ISO
228/1) female.
EXT60, EXT100, EXT160, EXT250, EXT600, EXT1000 pressure module pressure
connection G 1/4 (ISO 228/1) male.
Wetted parts AISI316 stainless steel, Hastelloy, Nitrile rubber.
In the PM20C, EXT20C, PM60, EXT60, PM100, EXT100,
PM160, EXT160, PM250 and EXT250 the maximum overpressure is twice the
range.
The maximum overpressure for EXT600 is 900 bar and for EXT1000 1100 bar.
HART® is a registered trademark of HART®
Communication Foundation.
Copyright © - Beamex 2005 |