MC5-IS is a calibrator that grows with the needs of the
user. MC5-IS can be used as a simple precision pressure indicator and
still easily be upgraded into a documenting multifunction pressure-temperature-electrical
calibrator, which communicates with calibration software.
| |
Display |
96 x 72 mm (3.78" x 2.83"), 320 x
240 pixels |
| |
Weight |
1.7 - 2.3 kg (3.7 - 5.1 lbs) |
| |
Dimensions |
245 mm (9.6") x 192 mm (7.5") x 74
mm (2.9"), (d/w/h) |
| |
Case protection |
IP65 (dust and water proof) |
| |
Keyboard |
Membrane protected individual keys |
| |
Battery type |
Easy to change rechargeable NiMH pack, 1200
mAh, 8.4 V DC |
| |
Battery operation |
Average 5 hours |
| |
Charger supply |
100
240 VAC, 50-60 Hz |
| |
Operating temperature |
-10
50 °C (14
122 °F) |
| |
Specifications valid |
15...35 °C (59...95 °F) |
| |
Temperature coefficient |
<±0.001% RDG / °C outside of
15...35 °C (59...95 °F) |
| |
Storage temperature |
-20 to 60 °C (-4 to 140 °F) |
| |
Humidity |
0 to 80% R.H. non condensing |
| |
Measurement sample rate |
2.5 / second |
| |
Warranty |
3 years as standard, battery pack 1 year |
NOTE: All specifications are subject to change without
prior notice.
MC5-IS is part of the Beamex's QCAL® System, which
means that your MC5-IS connects easily to a calibration system, which
typically comprises of portable calibrators, calibration software and
test benches.
Rechargeable NiMH battery pack
Battery Charger
Test Leads and Clips
Carrying Strap
RS232 Cable
Traceable Calibration Certificate
User's Manual
Pressure Hose Set (if MC5-IS is equipped with INT pressure modules)
Three Years Warranty
| |
| |
mV measurement 2) |
± 1000 mV |
0.001 - 0.01 mV |
0.02% RDG + 5 µV |
| |
V measurement 3) |
± 50 V |
0.00001 - 0.001 V |
0.02% RDG + 0.25 mV |
| |
mA measurement 4) |
± 100 mA |
0.0001 - 0.001 mA |
0.02% RDG + 1.5 µA |
| |
Hz measurement 5) |
0.0028 to 50000 Hz |
0.000001 - 0.1 Hz |
0.01% RDG |
| |
Pulse counting 5) |
0 to 9 999 999 pulses |
1 pulse |
N/A |
| |
mA generation 6) |
0 to 25 mA |
0.0001 mA |
0.02% RDG + 1.5 µA |
1) Uncertainty includes reference standard
uncertainty, hysteresis, nonlinearity, repeatability and typical long
term stability for mentioned period. (k=2)
2) Bias current <10 nA
3) Impedance >1 Mohm
4) Impedance < 7.5 ohm
5) Impedance > 1 Mohm. Frequency measurement minimum amplitude 0.5
Vpp (< 5 kHz),
1Vpp (5...50 kHz). Pulse counting minimum amplitude 0.5 Vpp (pulse length
> 100 µs),
1 Vpp (pulse length 100 µs...10 µs).
Trigger level range -1
+15 V.
6) Maximum load impedance 800 ohm
| |
INT B
EXT B |
kPa a
mbar a
psi a |
80 ... 120
800 ... 1200
11.6 ... 17.4 |
0.01
0.1
0.001 |
0.03 kPa
0.3
mbar
0.0044 psi
|
0.05 kPa
0.5 mbar
0.0073 psi |
| |
INT10mD
EXT10mD |
kPa
mbar
iwc |
± 1
± 10
± 4 |
0.0001
0.001
0.001 |
0.05 span |
0.05% span + 0.1% RDG |
| |
INT100m
EXT100m |
kPa
mbar
iwc |
0 to 10
0 to 100
0 to 40 |
0.0001
0.001
0.001 |
0.015% FS + 0.0125% RDG |
0.025% FS + 0.025% RDG |
| |
INT400mC
EXT400mC |
kPa
mbar
psi |
± 40
± 400
± 160 |
0.001
0.01
0.001 |
0.01% FS +
0.0125% RDG |
0.02% FS + 0.025%
RDG |
| |
INT1C
EXT1C |
kPa
bar
psi |
± 100
± 1
-14.5 to 15 |
0.001
0.00001
0.0001 |
0.007% FS + 0.0125% RDG |
0.015% FS + 0.025% RDG |
| |
INT2C
EXT2C |
kPa
bar
psi |
-100 to 200
-1 to 2
-14.5 to 30 |
0.001
0.00001
0.0001 |
0.005% FS + 0.01% RDG |
0.01% FS + 0.025% RDG |
| |
INT6C
EXT6C |
kPa
bar
psi |
-100 to 600
-1 to 6
-14.5 to 90 |
0.01
0.0001
0.001 |
0.005% FS + 0.01% RDG |
0.01% FS + 0.025% RDG |
| |
INT20C
EXT20C |
kPa
bar
psi |
-100 to 2000
-1 to 20
-14.5 to 300 |
0.01
0.0001
0.001 |
0.005% FS + 0.01% RDG |
0.01% FS + 0.025% RDG |
| |
INT60
EXT60 |
kPa
bar
psi |
0 to 6000
0 to 60
0 to 900 |
0.1
0.001
0.01 |
0.005% FS + 0.0125% RDG |
0.01% FS + 0.025% RDG |
| |
INT100
EXT100 |
MPa
bar
psi |
0 to 10
0 to 100
0 to 1500 |
0.0001
0.001
0.01 |
0.005% FS + 0.0125% RDG |
0.01% FS + 0.025% RDG |
| |
INT160
EXT160 |
MPa
bar
psi |
0 to 16
0 to 160
0 to 2400 |
0.0001
0.001
0.01 |
0.005% FS + 0.0125% RDG |
0.01% FS + 0.025% RDG |
| |
-
EXT250 |
MPa
bar
psi |
0 to 25
0 to 250
0 to 3700 |
0.001
0.01
0.1 |
0.007% FS + 0.0125% RDG |
0.015% FS + 0.025% RDG |
| |
-
EXT600 |
MPa
bar
psi |
0 to 60
0 to 600
0 to 9000 |
0.001
0.01
0.1 |
0.007% FS + 0.01% RDG |
0.015% FS + 0.025% RDG |
| |
-
EXT1000 |
MPa
bar
psi |
0 to 100
0 to 1000
0 to 15000 |
0.001
0.01
0.1 |
0.007% FS + 0.01% RDG |
0.015% FS + 0.025% RDG |
1) The MC5 / MC5-IS Calibrators can hold three internal pressure modules.
2) Every internal/external pressure module’s range may also be displayed in absolute pressure if the Barometric Module (B) is installed.
3) ‘Accuracy’ includes hysteresis, nonlinearity, repeatability and reference standard uncertainty (k=2).
4) ‘1 Year Uncertainty’ includes hysteresis, nonlinearity, repeatability and typical long-term stability for mentioned period (k=2).
All external pressure modules (EXT) are also compatible with Beamex MC2 and MC5P Calibrators.
Supports the following pressure units as standard:
Pa, hPa, kPa, MPa, mbar, bar, lbf/ft 2 , psi, gf/cm 2 , kgf/cm 2 , kgf/m
2 , kp/cm 2 , at, mmH 2 O, cmH 2 O, mH 2 O, iwc, ftH 2 O, mmHg, cmHg,
mHg, inHg, mmHg(0°C), inHg(0°C), mmH 2 O(4°C), inH 2 O(4°C),
ftH 2 O(4°C), inH 2 O(60°F), mmH 2 O(68°F), inH 2 O(68°F),
ftH 2 O(68°F), torr, atm.
Pressure modules 20bar and below, pressure connection
G 1/8 (ISO 228/1) 60° internal cone.
INT60-IS, INT100-IS, INT160-IS module pressure connection G 1/8 (ISO 228/1)
female.
EXT60-IS, EXT100-IS, EXT160-IS, EXT250-IS, EXT600-IS, EXT1000-IS pressure
module
pressure connection G 1/4 (ISO 228/1) male.
Wetted parts AISI316 stainless steel, Hastelloy, Nitrile rubber.
In the INT20C-IS, EXT20C-IS, INT60-IS, EXT60-IS,
INT100-IS, EXT100-IS, INT160-IS, EXT160-IS and EXT250-IS the maximum overpressure
is twice the range.
The maximum overpressure for EXT600-IS is 900 bar and for EXT1000-IS 1100
bar.
HART® is a registered trademark of HART®
Communication Foundation.
| |
mV generation 2) |
± 500 mV |
0.001 - 0.01 mV |
0.02% RDG + 4 µV |
| |
V generation 3) |
-2.5 to 10 V |
0.00001 - 0.0001 V |
0.02% RDG + 0.1 mV |
| |
mA sink |
0 to 25 mA |
0.0001 mA |
0.02% RDG + 1 µA |
| |
Hz generation
|
0.00028 to 50000 Hz |
0.000001 - 0.1 Hz |
0.01% RDG |
| |
Pulse generation |
0 to 9 999 999 pulses |
1 pulse |
N/A |
| |
Ohm simulation |
1 to 4000 ohm |
0.01 - 0.1 ohm |
0.04% RDG or 30 mohm 5) |
| |
Ohm measurement 4) |
0 to 4000 ohm |
0.001 - 0.1 ohm |
0.02% RDG + 3.5 mohm |
| |
mV measurement 6) |
± 250 mV |
0.001 - 0.01 mV |
0.02% RDG + 4 µV |
1) Uncertainty includes reference standard uncertainty,
hysteresis, nonlinearity,
repeatability and typical long term stability
for mentioned period. (k=2)
2) Load effect < 5 uV/mA. Maximum output current 1 mA.
3) Load effect < 100 uV/mA. Maximum output current 1 mA (0 .. 10 V).
4) Which ever is greater.
5) Specification valid with 4 wire connection. In 3 wire connection add
10 mohm.
6) Bias current < 10 nA
| |
Pt-sensors |
-200 to 850°C |
-200 to 0°C |
0.06°C
|
0.1°C |
| |
|
|
0 to 850°C |
0.025% RDG + 0.06°C |
0.025% RDG + 0.1°C |
Resolution 0.01°C.
| Pt50 (385) |
Pt500 (385) |
Pt100 (391) |
Ni100 (618) |
| Pt100 (385) |
Pt1000 (385) |
Pt100 (375) |
Ni120 (672) |
| Pt200 (385) |
Pt100 (3923) |
Pt100 (3926) |
Cu10 (427) |
| Pt400 (385) |
Pt100 (389) |
|
|
Also other RTD types available as option.
| |
B 2) |
0 to 1820 |
0 to 200
200 to 500
500 to 800
800 to 1820 |
3)
2.0°C
0.6°C
0.8°C |
| |
R 2) |
-50 to 1768 |
-50 to 0
0 to 150
150 to 1400
1400 to 1768 |
1.0°C
0.7°C
0.5°C
0.6°C |
| |
S 2) |
-50 to 1768 |
-50 to 0
0 to 50
50 to 1500
1500 to 1768 |
1.0°C
0.7°C
0.6°C
0.7°C |
| |
E 2) |
-270 to 1000 |
-270 to -200
-200 to 0
0 to 600
600 to 1000 |
3)
0.08% RDG + 0.07°C
0.015% RDG + 0.07°C
0.026% RDG |
| |
J 2) |
-210 to 1200 |
-210 to -200
-200 to 0
0 to 1200 |
3)
0.07%RDG + 0.08°C
0.02% RDG + 0.08°C |
| |
K 2) |
-270 to 1372 |
-270 to -200
-200 to 0
0 to 1000
1000 to 1372 |
3)
0.1% RDG + 0.1°C
0.02% RDG + 0.1°C
0.03% RDG |
| |
N 2) |
-270 to 1300 |
-270 to -200
-200 to -100
-100 to 0
0 to 750
750 to 1300 |
3)
0.2% RDG
0.05% RDG + 0.15°C
0.01% RDG + 0.15°C
0.03% RDG |
| |
T 2) |
-270 to 400 |
-270 to -250
-250 to -200
-200 to 0
0 to 400 |
3)
0.7°C
0.1% RDG + 0.1°C
0.01% RDG + 0.1°C |
| |
U 4) |
200 to 600 |
-200 to 0
0 to 600 |
0.1% RDG + 0.15°C
0.01% RDG + 0.15°C |
| |
L 4) |
-200 to 900 |
-200 to 0
0 to 900 |
0.07% RDG + 0.13°C
0.02% RDG + 0.13°C |
Resolution 0.01°C.
With internal reference junction (module RJ) add 0.1°C uncertainty.
Thermocouple types C 3) (ASTM E 988 - 96), G 3)
(ASTM E 1751 - 95e1) and D 3) (ASTM E 988 - 96) also available as standard.
Also other thermocouple types available as option.
1) Uncertainty includes reference standard
uncertainty, hysteresis, nonlinearity, repeatability and typical long
term stability for
mentioned period. (k=2)
2) IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
3) ±(0.02 % RDG + 4) mV
4) DIN 43710
Allows MC5-IS to communicate with Beamex's calibration software; QD3 documenting
software
and CMX calibration management software.
Allows MC5-IS to work as a datalogger. MC5-IS can log up to seven (7)
measurement channels simultaneously. The datalogging results can be downloaded
into a computer through the RS port for further examination. MC5-IS can
datalog more than 10000 measurement points.
Allows MC5-IS to communicate with HART® instruments by supporting
those HART® read/write commands typically used during the calibration
process.
The carrying case is made of durable, water and stain resistant fabric.
It provides room for test leads, pressure tubing, fittings and pressure
hand pumps. The MC5 fits within the internal padded compartment, which
includes room for the MC5 User Guide.
These lightweight, rugged hand pumps are ideal for the generation of pressure/vacuum.
Each hand pump comes with a variable volume adjustment knob and variable
vent knob for precise pressure/ vacuum control. Three pumps cover the
range from vacuum to 700 bar / 10,000 psi.
Beamex QCAL® calibration software automates the creation of instrument
calibration procedures which then can be downloaded to the MC5-IS. Depending
upon user requirements, either QD3 documenting software or CMX calibration
management software be used with MC5-IS to satisfy the requirements of
ISO 9000, QS-9000, OSHA, FDA , etc.
Copyright © - Beamex 2007 |