home > portable calibrators > MC5-IS > specifications

  MC5-IS:

  » introduction
  » features
  » specifications

  » downloads
  » related topics

 


 

Technical specifications:

General Specifications
Part of the System
Standard accessories (delivered with MC5-IS)
Electrical module (E)
Pressure Modules
Temperature Electrical Module (ET)
RTD Measurement and Simulation
Thermocouple Measurement and Simulation
Reference Junction Module (RJ)
MC5-IS Options
Optional Accessories

General Specifications

MC5-IS is a calibrator that grows with the needs of the user. MC5-IS can be used as a simple precision pressure indicator and still easily be upgraded into a documenting multifunction pressure-temperature-electrical calibrator, which communicates with calibration software.

- General
  Display 96 x 72 mm (3.78" x 2.83"), 320 x 240 pixels
  Weight 1.7 - 2.3 kg (3.7 - 5.1 lbs)
  Dimensions 245 mm (9.6") x 192 mm (7.5") x 74 mm (2.9"), (d/w/h)
  Case protection IP65 (dust and water proof)
  Keyboard Membrane protected individual keys
  Battery type Easy to change rechargeable NiMH pack, 1200 mAh, 8.4 V DC
  Battery operation Average 5 hours
  Charger supply 100…240 VAC, 50-60 Hz
  Operating temperature -10…50 °C (14…122 °F)
  Specifications valid 15...35 °C (59...95 °F)
  Temperature coefficient <±0.001% RDG / °C outside of 15...35 °C (59...95 °F)
  Storage temperature -20 to 60 °C (-4 to 140 °F)
  Humidity 0 to 80% R.H. non condensing
  Measurement sample rate 2.5 / second
  Warranty 3 years as standard, battery pack 1 year

NOTE: All specifications are subject to change without prior notice.

back to top


Part of System

MC5-IS is part of the Beamex's QCAL® System, which means that your MC5-IS connects easily to a calibration system, which typically comprises of portable calibrators, calibration software and test benches.

back to top


Standard Accessories (delivered with MC5-IS)

• Rechargeable NiMH battery pack
• Battery Charger
• Test Leads and Clips
• Carrying Strap
• RS232 Cable
• Traceable Calibration Certificate
• User's Manual
• Pressure Hose Set (if MC5-IS is equipped with INT pressure modules)
• Three Years Warranty

back to top


Electrical Module (E)

  Function Range Resolution 1 Year Uncertainty(±)1)
  mV measurement 2) ± 1000 mV 0.001 - 0.01 mV 0.02% RDG + 5 µV
  V measurement 3) ± 50 V 0.00001 - 0.001 V 0.02% RDG + 0.25 mV
  mA measurement 4) ± 100 mA 0.0001 - 0.001 mA 0.02% RDG + 1.5 µA
  Hz measurement 5) 0.0028 to 50000 Hz 0.000001 - 0.1 Hz 0.01% RDG
  Pulse counting 5) 0 to 9 999 999 pulses 1 pulse N/A
  mA generation 6) 0 to 25 mA 0.0001 mA 0.02% RDG + 1.5 µA

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned period. (k=2)
2) Bias current <10 nA
3) Impedance >1 Mohm
4) Impedance < 7.5 ohm
5) Impedance > 1 Mohm. Frequency measurement minimum amplitude 0.5 Vpp (< 5 kHz),
1Vpp (5...50 kHz). Pulse counting minimum amplitude 0.5 Vpp (pulse length > 100 µs),
1 Vpp (pulse length 100 µs...10 µs).
Trigger level range -1…+15 V.
6) Maximum load impedance 800 ohm

back to top


Internal (INT) and External (EXT) Pressure Modules

  Modules 1) Units Range 2) Resolution Accuracy 3)(±) 1 Year Uncertainty 4)(±)
 

INT B
EXT B

kPa a
mbar a
psi a
80 ... 120
800 ... 1200
11.6 ... 17.4
0.01
0.1
0.001
0.03 kPa
0.3 mbar
0.0044 psi
0.05 kPa
0.5 mbar
0.0073 psi
  INT10mD
EXT10mD
kPa
mbar
iwc
± 1
± 10
± 4
0.0001
0.001
0.001
0.05 span   0.05% span + 0.1% RDG
  INT100m
EXT100m
kPa
mbar
iwc
0 to 10
0 to 100
0 to 40
0.0001
0.001
0.001
0.015% FS + 0.0125% RDG 0.025% FS + 0.025% RDG
  INT400mC
EXT400mC
kPa
mbar
psi
± 40
± 400
± 160
0.001
0.01
0.001
0.01% FS + 0.0125% RDG 0.02% FS + 0.025% RDG
  INT1C
EXT1C
kPa
bar
psi
± 100
± 1
-14.5 to 15
0.001
0.00001
0.0001
0.007% FS + 0.0125% RDG 0.015% FS + 0.025% RDG
  INT2C
EXT2C
kPa
bar
psi
-100 to 200
-1 to 2
-14.5 to 30
0.001
0.00001
0.0001
0.005% FS + 0.01% RDG 0.01% FS + 0.025% RDG
  INT6C
EXT6C
kPa
bar
psi
-100 to 600
-1 to 6
-14.5 to 90
0.01
0.0001
0.001
0.005% FS + 0.01% RDG 0.01% FS + 0.025% RDG
  INT20C
EXT20C
kPa
bar
psi
-100 to 2000
-1 to 20
-14.5 to 300
0.01
0.0001
0.001
0.005% FS + 0.01% RDG 0.01% FS + 0.025% RDG
  INT60
EXT60
kPa
bar
psi
0 to 6000
0 to 60
0 to 900
0.1
0.001
0.01

0.005% FS + 0.0125% RDG

0.01% FS + 0.025% RDG
  INT100
EXT100
MPa
bar
psi
0 to 10
0 to 100
0 to 1500
0.0001
0.001
0.01
0.005% FS + 0.0125% RDG 0.01% FS + 0.025% RDG
  INT160
EXT160
MPa
bar
psi
0 to 16
0 to 160
0 to 2400
0.0001
0.001
0.01
0.005% FS + 0.0125% RDG 0.01% FS + 0.025% RDG
  -
EXT250
MPa
bar
psi
0 to 25
0 to 250
0 to 3700
0.001
0.01
0.1
0.007% FS + 0.0125% RDG 0.015% FS + 0.025% RDG
  -
EXT600
MPa
bar
psi
0 to 60
0 to 600
0 to 9000
0.001
0.01
0.1
0.007% FS + 0.01% RDG 0.015% FS + 0.025% RDG
  -
EXT1000
MPa
bar
psi
0 to 100
0 to 1000
0 to 15000
0.001
0.01
0.1
0.007% FS + 0.01% RDG 0.015% FS + 0.025% RDG

1) The MC5 / MC5-IS Calibrators can hold three internal pressure modules.
2) Every internal/external pressure module’s range may also be displayed in absolute pressure if the Barometric Module (B) is installed.
3) ‘Accuracy’ includes hysteresis, nonlinearity, repeatability and reference standard uncertainty (k=2).
4) ‘1 Year Uncertainty’ includes hysteresis, nonlinearity, repeatability and typical long-term stability for mentioned period (k=2).

All external pressure modules (EXT) are also compatible with Beamex MC2 and MC5P Calibrators.

Supports the following pressure units as standard:
Pa, hPa, kPa, MPa, mbar, bar, lbf/ft 2 , psi, gf/cm 2 , kgf/cm 2 , kgf/m 2 , kp/cm 2 , at, mmH 2 O, cmH 2 O, mH 2 O, iwc, ftH 2 O, mmHg, cmHg, mHg, inHg, mmHg(0°C), inHg(0°C), mmH 2 O(4°C), inH 2 O(4°C), ftH 2 O(4°C), inH 2 O(60°F), mmH 2 O(68°F), inH 2 O(68°F), ftH 2 O(68°F), torr, atm.

Pressure modules 20bar and below, pressure connection G 1/8 (ISO 228/1) 60° internal cone.
INT60-IS, INT100-IS, INT160-IS module pressure connection G 1/8 (ISO 228/1) female.
EXT60-IS, EXT100-IS, EXT160-IS, EXT250-IS, EXT600-IS, EXT1000-IS pressure module
pressure connection G 1/4 (ISO 228/1) male.
Wetted parts AISI316 stainless steel, Hastelloy, Nitrile rubber.

In the INT20C-IS, EXT20C-IS, INT60-IS, EXT60-IS, INT100-IS, EXT100-IS, INT160-IS, EXT160-IS and EXT250-IS the maximum overpressure is twice the range.
The maximum overpressure for EXT600-IS is 900 bar and for EXT1000-IS 1100 bar.

HART® is a registered trademark of HART® Communication Foundation.

back to top


Temperature Electrical Module (ET)

  Function Range Resolution 1 Year Uncertainty (±)1
  mV generation 2) ± 500 mV 0.001 - 0.01 mV 0.02% RDG + 4 µV
  V generation 3) -2.5 to 10 V 0.00001 - 0.0001 V 0.02% RDG + 0.1 mV
  mA sink 0 to 25 mA 0.0001 mA 0.02% RDG + 1 µA
  Hz generation
0.00028 to 50000 Hz 0.000001 - 0.1 Hz 0.01% RDG
  Pulse generation 0 to 9 999 999 pulses 1 pulse N/A
  Ohm simulation 1 to 4000 ohm 0.01 - 0.1 ohm 0.04% RDG or 30 mohm 5)
  Ohm measurement 4) 0 to 4000 ohm 0.001 - 0.1 ohm 0.02% RDG + 3.5 mohm
  mV measurement 6) ± 250 mV 0.001 - 0.01 mV 0.02% RDG + 4 µV

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity,
    repeatability and typical long term stability for mentioned period. (k=2)
2) Load effect < 5 uV/mA. Maximum output current 1 mA.
3) Load effect < 100 uV/mA. Maximum output current 1 mA (0 .. 10 V).
4) Which ever is greater.
5) Specification valid with 4 wire connection. In 3 wire connection add 10 mohm.
6) Bias current < 10 nA

back to top


RTD Measurement and Simulation

  Type Range (°C) Range (°C) Measurement
1 Year Uncert. 1) (±)
Simulation
1 Year Uncert. 1) (±)
  Pt-sensors -200 to 850°C -200 to 0°C 0.06°C
0.1°C
      0 to 850°C 0.025% RDG + 0.06°C 0.025% RDG + 0.1°C

Resolution 0.01°C.

Other RTD types available as standard
• Pt50 (385) • Pt500 (385) • Pt100 (391) • Ni100 (618)
• Pt100 (385) • Pt1000 (385) • Pt100 (375) • Ni120 (672)
• Pt200 (385) • Pt100 (3923) • Pt100 (3926) • Cu10 (427)
• Pt400 (385) • Pt100 (389)    

Also other RTD types available as option.

back to top


Thermocouple Measurement and Simulation

  Type Range (°C) Range (°C) 1 Year Uncertainty (±)
  B 2) 0 to 1820 0 to 200
200 to 500
500 to 800
800 to 1820
3)
2.0°C
0.6°C
0.8°C
  R 2) -50 to 1768 -50 to 0
0 to 150
150 to 1400
1400 to 1768
1.0°C
0.7°C
0.5°C
0.6°C
  S 2) -50 to 1768 -50 to 0
0 to 50
50 to 1500
1500 to 1768
1.0°C
0.7°C
0.6°C
0.7°C
  E 2) -270 to 1000 -270 to -200
-200 to 0
0 to 600
600 to 1000
3)
0.08% RDG + 0.07°C
0.015% RDG + 0.07°C
0.026% RDG
  J 2) -210 to 1200 -210 to -200
-200 to 0
0 to 1200
3)
0.07%RDG + 0.08°C
0.02% RDG + 0.08°C
  K 2) -270 to 1372 -270 to -200
-200 to 0
0 to 1000
1000 to 1372
3)
0.1% RDG + 0.1°C
0.02% RDG + 0.1°C
0.03% RDG
  N 2) -270 to 1300 -270 to -200
-200 to -100
-100 to 0
0 to 750
750 to 1300
3)
0.2% RDG
0.05% RDG + 0.15°C
0.01% RDG + 0.15°C
0.03% RDG
  T 2) -270 to 400 -270 to -250
-250 to -200
-200 to 0
0 to 400
3)
0.7°C
0.1% RDG + 0.1°C
0.01% RDG + 0.1°C
  U 4) 200 to 600 -200 to 0
0 to 600
0.1% RDG + 0.15°C
0.01% RDG + 0.15°C
  L 4) -200 to 900 -200 to 0
0 to 900
0.07% RDG + 0.13°C
0.02% RDG + 0.13°C

Resolution 0.01°C.
With internal reference junction (module RJ) add 0.1°C uncertainty.

Thermocouple types C 3) (ASTM E 988 - 96), G 3) (ASTM E 1751 - 95e1) and D 3) (ASTM E 988 - 96) also available as standard.
Also other thermocouple types available as option.

1) Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for
mentioned period. (k=2)
2) IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
3) ±(0.02 % RDG + 4) mV
4) DIN 43710

back to top


Reference Junction Module (RJ)

  Ambient temperature range 1 Year uncertainty(±) 1)
  -10 .. 50°C 0.1°C

back to top


MC5-IS OPTIONS

QCAL® Communication
Allows MC5-IS to communicate with Beamex's calibration software; QD3 documenting software
and CMX calibration management software.

Multichannel Datalogging
Allows MC5-IS to work as a datalogger. MC5-IS can log up to seven (7) measurement channels simultaneously. The datalogging results can be downloaded into a computer through the RS port for further examination. MC5-IS can datalog more than 10000 measurement points.

HART® Communication
Allows MC5-IS to communicate with HART® instruments by supporting those HART® read/write commands typically used during the calibration process.

back to top


OPTIONAL ACCESSORIES
MC5 is supported by a large number of accessories such as:

Carrying Case (soft)
The carrying case is made of durable, water and stain resistant fabric. It provides room for test leads, pressure tubing, fittings and pressure hand pumps. The MC5 fits within the internal padded compartment, which includes room for the MC5 User Guide.

Calibration Hand Pumps
These lightweight, rugged hand pumps are ideal for the generation of pressure/vacuum. Each hand pump comes with a variable volume adjustment knob and variable vent knob for precise pressure/ vacuum control. Three pumps cover the range from vacuum to 700 bar / 10,000 psi.

Calibration Software
Beamex QCAL® calibration software automates the creation of instrument calibration procedures which then can be downloaded to the MC5-IS. Depending upon user requirements, either QD3 documenting software or CMX calibration management software be used with MC5-IS to satisfy the requirements of ISO 9000, QS-9000, OSHA, FDA , etc.

back to top


Copyright © - Beamex 2007