Display |
96 x 72 mm (3.78" x 2.83"), 320 x
240 pixels, back lit LCD |
| |
Weight |
1.7 - 2.3 kg (3.7 - 5.1 lbs) |
| |
Dimensions |
245 mm (9.6") x 192 mm (7.5") x 74
mm (2.9"), (d/w/h) |
| |
Case protection |
IP65 (dust and water proof) |
| |
Keyboard |
Membrane protected individual keys |
| |
Battery type |
Easy to change rechargeable NiMH pack, 4000
mAh, 7.2 V DC |
| |
Battery operation |
Average 10hours |
| |
Charger supply |
100
240 VAC, 50-60 Hz |
| |
Operating temperature |
-10
50 °C (14
122 °F) |
| |
Specifications valid |
15...35 °C (59...95 °F) |
| |
Temperature coefficient |
<±0.001% RDG / °C outside of
15...35 °C (59...95 °F) |
| |
Storage temperature |
-20 to 60 °C (-4 to 140 °F) |
| |
Warranty |
2 years |
Rechargeable NiMH battery pack
Battery Charger
Test Leads and Clips
Carrying Strap
Calibration Certificate
Pressure Hose Set (if MC3 is equipped with the pressure section)
User 's Manual
| |
mV measurement |
± 500 mV |
0.001 - 0.01 mV |
0.02% RDG + 8 µV |
| |
mV generation |
± 500 mV |
0.001 - 0.01 mV |
0.02% RDG + 8 µV |
| |
V generation |
± 12 V |
0.00001 - 0.0001 V |
0.02% RDG + 0.2 mV |
| |
mA generation |
± 25 mA |
0.0001 mA |
0.02% RDG + 2 µA |
| |
Hz generation |
0.00028 to 50000 Hz |
0.000001 - 0.1 Hz |
0.02% RDG |
| |
Pulse generation |
0 to 9 999 999 pulses |
1 pulse |
N/A |
| |
Ohm measurement |
0 to 4000 ohm |
0.001 - 0.1 ohm |
0.02% RDG + 7 mohm |
| |
Ohm simulation |
1 to 4000 ohm |
0.01 - 0.1 ohm |
0.05% RDG or 50 mohm |
| |
Pt50 (385) |
Pt400 (385) |
Pt100 (3923) |
Pt100 (375) |
Ni120 (672) |
| |
Pt100 (385) |
Pt500 (385) |
Pt100 (389) |
Pt100 (3926) |
Cu10 (427) |
| |
Pt200 (385) |
Pt1000 (385) |
Pt100 (391) |
Ni100 (618) |
|
1 year uncertainty (±), measurement 1): 0.2°C.
1 year uncertainty (±), simulation 1): 0.25°C.
Resolution 0.01°C. With internal reference junction add 0.2°C
uncertainty.
1) Midrange uncertainty.
| |
B |
0 to 1820 |
1.0°C |
| |
R |
-50 to 1768 |
1.0°C |
| |
S |
-50 to 1768 |
1.0°C |
| |
E |
-270 to 1000 |
1.0°C |
| |
J |
-210 to 1200 |
0.25°C |
| |
K |
-270 to 1372 |
0.3 °C |
| |
N |
-270 to 1300 |
0.35°C |
| |
T |
-270 to 400 |
0.25°C |
| |
U |
-200 to 600 |
0.35°C |
| |
L |
-200 to 900 |
0.35°C |
Midrange uncertainty. Resolution 0.01°C. With
internal reference junction add 0.2°C uncertainty.
| |
PR2C 1) |
kPa
bar
psi |
-100 to 200
-1 to 2
-14.5 to 30 |
0.01
0.0001
0.001 |
0.05% FS |
| |
PR20C 1) |
kPa
bar
psi |
-100 to 2000
-1 to 20
-14.5 to 300 |
0.1
0.001
0.01
|
0.05% FS |
| |
PR160 1) |
MPa
bar
psi |
0 to 16
0 to 160
0 to 2400 |
0.001
0.01
0.1 |
0.05% FS |
| |
Barometric option 2) |
enables also absolute pressure
measurement for above pressure inputs. With barometic option add 0.1
kPa (0.0146 psi) uncertainty for absolute pressure measurement. |
1) Select one of the pressure ranges above (PR2C,
PR20C or PR160).
2) The Barometric option can be installed if one pressure range out of
PR2C, PR20C or PR160 is installed.
| |
V measurement |
± 50 V |
0.00001 - 0.001 V |
0.02% RDG + 0.5 mV |
| |
mA measurement |
± 100 mA |
0.0001 - 0.001 mA |
0.02% RDG + 2 µA |
| |
mA generation |
0 to 25 mA |
0.0001 mA |
0.02% RDG + 2 µA |
| |
Hz measurement |
0.0028 to 50000 Hz |
0.000001 - 0.1 Hz |
0.02% RDG |
| |
Pulse counting |
0 to 9 999 999 pulses |
1 pulse |
N/A |
Uncertainty for all sections above includes reference standard uncertainty,
hysteresis, nonlinearity, repeatability and typical long-term stability
for mentioned period. (k=2)
This option expands the measurement capability of the pressure section
to also include absolute pressure.
QCAL® Communication allows MC3 to communicate with Beamex 's calibration
software; QD3 documenting software or CMX calibration management software.
PG series of calibration pumps
Carrying case
Calibration Software