Display |
60 mm x 60 mm (2.36" x 2.36"),
160 x 160 pixels, back lit LCD |
| |
Weight |
770 g (1.7 lbs) |
| |
Dimensions |
215 mm (8.5") x 102 mm (4")
x 49 mm (1.9") (d/w/h) |
| |
Keyboard |
Membrane keyboard |
| |
Battery type |
Rechargeable NiMH pack, 4000 mAh,
3.6V DC |
| |
Charging time |
5 hours |
| |
Charger supply |
100 ... 240 VAC, 50-60 Hz |
| |
Battery operation |
13 ... 24 hours
in measurement mode, back light off.
8 ... 12 hours when taking 12 mA from loop supply, back light on. |
| |
Battery operation with
optional dry battery cartridge
and 4 alkaline AA cells |
4 ... 8 hours in measurement mode,
back light off.
3... 4 hours when taking 12 mA from loop supply, back light on. |
| |
Operating temperature |
-10 ... 50°C
(14 ... 122°F) |
| |
Operating temp. while charging
the batteries |
0 ... 35°C (32 ... 95°C) |
| |
Storage temperature |
-20 ... 60°C (-4 ... 140°F) |
| |
Humidity |
0 to 80% R.H. non condensing |
| |
Warmup time |
Specifications valid after a 5-minute
warmup period. |
| |
Max. input voltage |
30 V AC, 60 V DC |
| |
Safety |
Directive 73/23/EEC, EN 61010-1 |
| |
EMC |
Directive 89/336/EEC, EN 61326 |
| |
± 0.25 V |
0.001mV |
0.02% RDG + 5 µV |
| |
±(0.25 ... 1 V) |
0.01 mV |
0.02% RDG + 5 µV |
| |
1 ... 25 V |
0.1 mV |
0.02% RDG + 0.25 mV |
| |
25 ... 60 V |
1 mV |
0.02% RDG + 0.25 mV |
| |
Temperature coefficient |
< ±0.0015% RDG / °C
outside of 18 ... 28°C
< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F |
| |
Input impedance |
>1 Mohm |
| |
Supported units |
V, mV, µV |
| |
Display update rate |
3 / second |
| |
±25mA |
0.0001 mA |
0.02% RDG + 1.5 µA |
| |
±(25 ... 100 mA) |
0.001 mA |
0.02% RDG + 1.5 µA |
| |
Temperature coefficient |
< ±0.0015% RDG / °C
outside of 18 ... 28°C
< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F |
| |
Input impedance |
< 7.5 ohm |
| |
Supported units |
mA, µA |
| |
Display update rate |
3 / second |
| |
Output Current |
max 25 mA, short circuit protected |
| |
Output Voltage |
24 V ± 10% |
| |
Output impedance in HART® compatible
mode |
300 ohm ± 20% |
| |
0.0027 ... 0.5 Hz |
0.000001 Hz |
0.01% RDG |
| |
0.5 ... 5 Hz |
0.00001 Hz |
0.01% RDG |
| |
5 ... 50 Hz |
0.0001 Hz |
0.01% RDG |
| |
50 ... 500 Hz |
0.001 Hz |
0.01% RDG |
| |
500 ... 5000 Hz |
0.01 Hz |
0.01% RDG |
| |
5000 ... 50000 Hz |
0.1 Hz |
0.01% RDG |
| |
Temperature coefficient |
Specification valid from -10 to
50°C (14 ... 122°F) |
| |
Input impedance |
> 1 Mohm |
| |
Trigger level |
-1
14 V in 1 V steps and open
collector inputs |
| |
Minimum signal amplitude |
2 Vpp (< 10 kHz), 3 Vpp (10...50
kHz) |
| |
Supported units |
Hz, kHz, cph, cpm, 1/Hz (s), 1/kHz
(ms), 1/MHz (µs) |
| |
Gate period |
267 ms + 1 signal period |
1) Uncertainty includes reference standard uncertainty,
hysteresis, nonlinearity, repeatability
and typical long-term stability for mentioned period. (k=2)
| |
Range |
0 to 9 999 999 pulses |
| |
Input impedance |
> 1 Mohm |
| |
Trigger level |
-1
14 V in 1 V steps and
open collector inputs |
| |
Minimum signal amplitude |
2 Vpp (pulse length > 50 µs),
3 Vpp (pulse length 10 ... 50 µs) |
| |
Potential free contacts |
Test Voltage (Current) |
2.8 V (0.13 mA) or 24 V (35 mA) |
| |
Voltage level detection |
Input impedance |
> 1 Mohm |
| |
|
Trigger level |
-1
14 V in 1 V steps |
| |
IPM200mC |
kPa
mbar
iwc |
±20
±200
±80 |
0.001
0.01
0.01 |
0.05% RDG + 0.05% FS |
| |
IPM2C |
kPa
bar
psi |
-100 to 200
-1 to 2
-14.5 to 30 |
0.01
0.0001
0.001 |
0.05% FS |
| |
IPM20C |
kPa
bar
psi |
-100 to 2000
-1 to 20
-14.5 to 300 |
0.1
0.001
0.01 |
0.05% FS |
| |
IPM160 |
MPa
bar
psi |
0 to 16
0 to 160
0 to 2400 |
0.001
0.01
0.1 |
0.05% FS |
| |
Barometric option |
Enables also absolute pressure
measurement for above pressure inputs.
With barometric option add 0.1 kPa (0.0146 psi) uncertainty for absolute
pressure measurement. |
| |
Temperature coefficient |
< ±0.001 % RDG /°C
outside 15
35 °C.
< ±0.0006 % RDG /°F outside 59
95 °F |
| |
Maximum overpressure |
2 × Range |
| |
Pressure port |
G 1/8" female (G 1/8 (ISO
228/1) 60° internal cone adaptor) |
| |
Media compatibility |
Wetted parts: AISI316 stainless
steel, Nitrile rubber. |
| |
Supported pressure units |
Pa, hPa, kPa, MPa, mbar, bar, lbf/ft2,
psi, ozf/in2, gf/cm2, kgf/cm2, kgf/m2, kp/cm2, at, mmH2O, cmH2O, mH2O,
iwc, ftH2O, mmHg, cmHg, mHg, inHg, mmHg(0°C), inHg(0°C), mmH2O(4°C;
60°F; 68°F/20°C), cmH2O(4°C; 60°F; 68°F/20°C),
inH2O(4°C; 60°F; 68°F/20°C), ftH2O(4°C; 60°F;
68°F/20°C), torr, atm, + four (4) user configurable units |
| |
Display update rate |
2.5 / second |
| |
Barometric |
800
1200 mbar abs |
23.6
35.4 inHg a |
0.5 mbar (0.015 inHg) |
| |
EXT10mD |
±10 mbar
differential |
±4 iwc
differential |
0.1% RDG + 0.05%
Span |
| |
EXT100m |
0
100 mbar gauge |
0
40 iwc |
0.04% RDG + 0.025% FS |
| |
EXT400mC |
±400 mbar |
±160 iwc |
0.04% RDG + 0.02%
FS |
| |
EXT1C |
±1 bar |
-14.5
15 psi |
0.04% RDG + 0.01% FS |
| |
EXT2C |
-1
2 bar |
-14.5
30 psi |
0.04% RDG + 0.01%
FS |
| |
EXT6C |
-1
6 bar |
-14.5
90 psi |
0.04% RDG + 0.01% FS |
| |
EXT20C |
-1
20
bar |
-14.5
300 psi |
0.04% RDG + 0.01%
FS |
| |
EXT60 |
0
60 bar |
0
900 psi |
0.04% RDG + 0.01% FS |
| |
EXT100 |
0
100
bar |
0
1500
psi |
0.04% RDG + 0.01%
FS |
| |
EXT160
|
0
160 bar |
0
2400 psi |
0.04% RDG + 0.013% FS |
| |
EXT250 |
0
250
bar |
0
3700
psi |
0.04% RDG + 0.015%
FS |
| |
EXT600 |
0
600 bar |
0
9000 psi |
0.04% RDG + 0.015% FS |
| |
EXT1000 |
0
1000
bar |
0
15000
psi |
0.04% RDG + 0.015%
FS |
| |
EXT200mC-s |
±200 mbar
±80 iwc |
0.01 mbar
0.01 iwc |
0.05% RDG + 0.05% FS |
| |
EXT2C-s |
-1 ... 2 bar
-14.5 ... 30 psi |
0.0001 bar
0.001 psi |
0.05% FS |
| |
EXT20C-s |
-1 ... 20 bar
-14.5 ... 300 psi |
0.001 bar
0.01 psi |
0.05% FS |
| |
EXT160-s |
0 ... 160 bar
0 ... 2400 psi |
0.01 bar
0.1 psi |
0.05% FS |
1) Uncertainty includes reference standard
uncertainty, hysteresis, non-linearity, repeatability and typical long-term
stability for mentioned period. (k=2)
2) The internal pressure module's range may also be displayed in absolute
pressure if an external Barometric Module (EXT B) is used.
All external pressure modules are also compatible with Beamex MC5 and
MC5P Calibrators.
| |
-25 ... 150 mV |
0.001 mV |
0.02% RDG + 4 µV |
| |
Temperature coefficient |
< ±0.0015% RDG / °C
outside of 18 ... 28°C
< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F |
| |
Input impedance |
>10 Mohm |
| |
Supported units |
V, mV, µV |
| |
Display update rate |
3 / second |
| |
-25 ... 150 mV |
0.001 mV |
0.02% RDG + 4 µV |
| |
Temperature coefficient |
< ±0.0015% RDG / °C
outside of 18 ... 28°C
< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F |
| |
Maximum load current |
5 mA |
| |
Load effect |
<50 µV/mA |
| |
Supported units |
V, mV, µV |
| |
±0.25 V |
0.01 mV |
0.02% RDG + 0.1 mV |
| |
-3 ... -0.25 V |
0.1 mV |
0.02% RDG + 0.1 mV |
| |
0.25 ... 12 V |
0.1 mV |
0.02% RDG + 0.1 mV |
| |
Temperature coefficient |
< ±0.0015% RDG / °C
outside of 18 ... 28°C
< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F |
| |
Maximum load current |
5 mA |
| |
Load effect |
<50 µV/mA |
| |
Supported units |
V, mV, µV |
| |
0 ... 25 mA |
0.0001 mA |
0.02% RDG + 1.5 µV |
| |
Temperature coefficient |
< ±0.0015% RDG / °C
outside of 18 ... 28°C
< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F |
| |
Max. load impedance (source) |
750 ohm (0 ... 20 mA), 600 ohm
(20 ... 25 mA) |
| |
Max. loop voltage (sink) |
60 V |
| |
Supported units |
mA, µA |
| |
0 ... 250 ohm
250 ... 2650 ohm
2650 ... 4000 phm |
1 mohm
10 mohm
100 mohm |
4-wire connection:
0.02% RDG + 3.5 mohm
3-wire connection:
0.02% RDG + 13.5 mohm |
| |
Temperature coefficient |
< ±0.0015% RDG / °C
outside of 18 ... 28°C
< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F |
| |
Measurement current |
Pulsed, bi-directional 1 mA (0..500
W), 0.2 mA (>500 W). |
| |
Supported units |
ohm, kohm |
| |
Display update rate |
3 / second |
| |
0 ... 400 ohm |
10 mohm |
0.04% RDG or 30 mohm
(whichever is greater) |
| |
400 ... 4000 ohm |
100 mohm |
0.04% RDG or 30 mohm
(whichever is greater) |
| |
Temperature coefficient |
< ±0.0015% RDG / °C
outside of 18 ... 28°C
< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F |
| |
Max. resistance excitation current |
5 mA (0 ... 650 ohm)
Iexc
× Rsim < 3.25 v (650 ... 4000 ohm) |
| |
Setting time (pulsed currents) |
1 ms |
| |
Supported units |
ohm, kohm |
| |
0.0005 ... 0.5 Hz |
0.000001 Hz |
0.01% RDG |
| |
0.5 ... 5 Hz |
0.00001 Hz |
0.01% RDG |
| |
5 ... 50 Hz |
0.0001 Hz |
0.01% RDG |
| |
50 ... 500 Hz |
0.001 Hz |
0.01% RDG |
| |
500 ... 5000 Hz |
0.01 Hz |
0.01% RDG |
| |
5000 ... 10000 Hz |
0.1 Hz |
0.01% RDG |
| |
Temperature coefficient |
Specification valid from -10 to
50°C (14 ... 122°F) |
| |
Max. load current |
5 mA |
| |
Output amplitude positive square
wave |
0 ... 12 Vpp ±(0.2 V + 5%) |
| |
Output amplitude
symmetric square wave |
0 ... 6Vpp ±(0.2
V + 5%) |
| |
Duty cycle |
1 ... 99% (0.0009 ... 500 Hz),
high / low time: min 25 µs, max 1165 s |
| |
Supported units |
Hz, kHz, cph,
1/Hz (s), 1/kHz (ms), 1/MH (µs) |
| |
Jitter |
<0.28 µs |
| |
Range |
0 to 9 999 999 pulses |
| |
Resolution |
1 pulse |
| |
Max. load current |
5 mA |
| |
Output amplitude
positive pulse |
0 ... 12 Vpp
±(0.2 V + 5%) |
| |
Output amplitude symmetric pulse |
0 ... 6Vpp ±(0.2 V + 5%) |
| |
Pulse frequency |
0.0005 ... 10
000 Hz |
| |
Duty cycle |
1 ... 99 % (0.0009
500 Hz),
high / low time: min 25µs, max 1165 s |
| |
B 2) |
0 to 1820 |
0 to 200
200 to 400
400 to 1820 |
(3
2.0°C
1.0°C |
| |
R 2) |
-50 to 1768 |
-50 to 0
0 to 100
100 to 1768 |
1.0°C
0.8°C
0.6°C |
| |
S 2) |
-50 to 1768 |
-50 to 0
0 to 1768 |
1.0°C
0.7°C |
| |
E 2) |
-270 to 1000 |
-270 to -200
-200 to 1000 |
(3
0.25°C |
| |
J 2) |
-210 to 1200 |
-210 to 1200 |
0.3°C |
| |
K 2) |
-270 to 1372 |
-270 to -200
-200 to 1000
1000 to 1372 |
(3
0.3°C
0.4°C |
| |
N 2) |
-270 to 1300 |
-270 to -200
-200 to 1300 |
(3
0.4°C |
| |
T 2) |
-270 to 400 |
-270 to -200
-200 to -100
-100 to 400 |
(3
0.3°C
0.2°C |
| |
U 4) |
200 to 600 |
-200 to -100
-100 to 600 |
0.3°C
0.2°C |
| |
L 4) |
-200 to 900 |
-200 to 900 |
0.25°C |
| |
C 5) |
0 ... 2315 |
0 ... 1000
1000 ... 2000
2000 ... 2315 |
0.4°C
0.8°C
1.2°C |
| |
G 6) |
0 ... 2315 |
0 ... 100
100 ... 2315 |
(3
1.0°C |
| |
D 5) |
0 ... 2315 |
0 ... 1000
1000 ... 2000
2000 ... 2315 |
0.4°C
0.8°C
1.2°C |
| |
Resolution |
0.01°C |
0.01°C |
| |
Temperature coefficient |
< ±0.0015%
of thermovoltage / °C outside of 18 ... 28°C
< ±0.0008%
of thermovoltage / °F outside of 64.4 ... 82.4°F |
< ±0.0015%
of thermovoltage / °C outside of 18 ... 28°C
< ±0.0008%
of thermovoltage / °F outside of 64.4 ... 82.4°F |
| |
Input impedance |
>10 Mohm |
- |
| |
Supported units |
°C, °F,
K |
°C, °F,
K |
| |
Display update range |
3 / second |
- |
| |
Max. load current |
- |
5 mA |
| |
Load effect |
- |
< 5 µV/mA |
1) Uncertainty includes reference standard uncertainty,
hysteresis, non-linearity, repeatability and typical long-term stability
for mentioned period. (k=2).
Uncertainty does not include reference junction uncertainty.
2) IEC 584, NIST MN 175, BS 4937, ANSI MC96.1
3) ±0.02 % of thermovoltage + 4 µV
4) DIN 43710
5) ASTM E 988 - 96
6) ASTM E 1751 - 95e1
| |
Pt 50 ... 1000 |
-200 ... 200°C
200 ... 600°C
600 ... 850°C |
0.01°C
0.01°C
0.01°C |
0.1°C
0.2°C
0.3°C |
| |
Ni 100 |
-60 ... 180°C |
0.01°C |
0.1°C |
| |
Ni 120 |
-80 ... 260°C |
0.01°C |
0.1°C |
| |
Cu 10 |
-200 ... 260°C |
0.01°C |
0.2°C |
| |
Temperature coefficient |
< ±0.0015% RDG / °C
outside of 18 ... 28°C
< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F |
| |
Measurement current |
Pulsed, 1 mA (0 ... 500 ohm), 0.2
mA (>500 ohm) |
| |
Supported units |
°C, °F, K |
| |
Display update rate |
3 / second |
| |
Pt 50 ... 1000 |
-200 ... 200°C
200 ... 600°C
600 ... 850°C |
0.01°C
0.01°C
0.01°C |
0.15°C
0.25°C
0.35°C |
| |
Ni 100 |
-60 ... 180°C |
0.01°C |
0.15°C |
| |
Ni 120 |
-80 ... 260°C |
0.01°C |
0.15°C |
| |
Cu 10 |
-200 ... 260°C |
0.01°C |
0.8°C |
| |
Temperature coefficient |
< ±0.0015% RDG / °C
outside of 18 ... 28°C
< ±0.0008% RDG / °F outside of 64.4 ... 82.4°F |
| |
Max. resistance excitation current |
5 mA (0 ... 650 ohm)
Iexc × Rsim < 3.25 V (680 ... 4000 ohm) |
| |
Supported units |
°C, °F, K |
| |
Pt 50 (385) |
Pt 500 (385) |
Pt 100 (3926) |
Ni 100 (618) |
| |
Pt 100 (385) |
Pt 1000 (385) |
Pt 100 (391) |
Ni 120 (672) |
| |
Pt 200 (385) |
|
Pt 100 (375) |
Cu 10 (427) |
| |
Pt 400 (385) |
|
Pt 100 (3923) |
|
User guide
Calibration certificate
Internal rechargeable NiMH battery pack + battery charger
A pair of test leads and clips
USB cable
Pressure connector adapter from G1/8" female to G 1/8"
male with 60° internal cone (included in PE and MF models)
Pressure T-hose
Soft carrying case
Connection cable for external (EXT) pressure modules
Dry battery cartridge
Calibration handpumps
Copyright © - Beamex 2008 |